SAFETY INFORMATION<br />

8<br />

tubes, the bottles and tubes should contain no<br />

more than 3/4 the total volume of the container.<br />

Chamber should be cleaned daily when processing<br />

media.<br />

19. Keep patients at least six (6) feet away from the<br />

sterilizer when it is in use.<br />

20. Do not operate this sterilizer in the presence of<br />

flammable or explosive substances (such as anesthetics).<br />

A fire or explosion could occur.<br />

21. Never use a wire brush, steel wool, abrasive material,<br />

or chloride-containing products to clean door<br />

and chamber assembly.<br />

22. Use a process monitor in every cycle to ensure<br />

exposure to minimum sterilizing conditions. Use a<br />

biological indicator once a week to verify your sterilization<br />

process.<br />

23. Dental handpieces should be run only on optional<br />

cycle at 132°C for 10 minutes. Check manufacturer's<br />

recommendations for sterilization parameters.