STERILIZATION PROCEDURES<br />

32<br />

Warning<br />

<strong>Sterilizer</strong> and trays will be hot<br />

after cycle is run. Always wear<br />

protective gloves and apron when<br />

removing a processed load.<br />

Protective gloves and apron<br />

should also be worn when reloading<br />

sterilizer following previous<br />

operation.<br />

Open door slowly at the end of a<br />

liquid sterilization cycle. Do not<br />

allow hot bottles to be jolted. This<br />

can cause bottle explosions. Do<br />

not move bottles if any boiling or<br />

bubbling is present.<br />

Allow bottles to cool before<br />

attempting to move them from<br />

sterilizer shelf or tray(s) to the<br />

storage area.<br />

Handling Sterilized Materials<br />

Loads may be removed from the sterilizer at the end of the<br />

drying cycle. (Liquids may be removed at the end of the sterilization<br />

cycle.) Unwrapped items should be used immediately.<br />

Any unwrapped items not used immediately cannot be considered<br />

sterile after removal from the sterilizer chamber.<br />

Date all wrapped packages and packs prior to storage.<br />

Check the integrity of all wrapping materials prior to opening<br />

and using wrapped or packed items; if the wrapping materials<br />

are opened or torn, the instruments must be considered<br />

potentially unsterile and should be re-sterilized prior to use.<br />

Re-sterilize all wrapped packages and packs after one month<br />

in storage.