STERILIZATION PROCEDURES<br />
32<br />
Warning<br />
<strong>Sterilizer</strong> and trays will be hot<br />
after cycle is run. Always wear<br />
protective gloves and apron when<br />
removing a processed load.<br />
Protective gloves and apron<br />
should also be worn when reloading<br />
sterilizer following previous<br />
operation.<br />
Open door slowly at the end of a<br />
liquid sterilization cycle. Do not<br />
allow hot bottles to be jolted. This<br />
can cause bottle explosions. Do<br />
not move bottles if any boiling or<br />
bubbling is present.<br />
Allow bottles to cool before<br />
attempting to move them from<br />
sterilizer shelf or tray(s) to the<br />
storage area.<br />
Handling Sterilized Materials<br />
Loads may be removed from the sterilizer at the end of the<br />
drying cycle. (Liquids may be removed at the end of the sterilization<br />
cycle.) Unwrapped items should be used immediately.<br />
Any unwrapped items not used immediately cannot be considered<br />
sterile after removal from the sterilizer chamber.<br />
Date all wrapped packages and packs prior to storage.<br />
Check the integrity of all wrapping materials prior to opening<br />
and using wrapped or packed items; if the wrapping materials<br />
are opened or torn, the instruments must be considered<br />
potentially unsterile and should be re-sterilized prior to use.<br />
Re-sterilize all wrapped packages and packs after one month<br />
in storage.