63<br />
Host Interface<br />
containing one ASCII character for each qualification that is<br />
asserted. If the qualification character is present, the qualification is<br />
asserted. If the qualification condition character is absent, the qualification<br />
is not asserted. Each character present has a unique<br />
meaning, as described in the following table.<br />
Table 9. Flags Character Definitions<br />
QUALIFICATION CHARACTER QUALIFICATION MEANING<br />
R Over-range error<br />
N Negative power<br />
S Measurement is sped up<br />
T Over-temperature error<br />
0 (zero) No qualification exists<br />
Note that X and Y positions are zero if the power is less than 10% of<br />
the minimum range.<br />
Note that the data has over-range status if the power is greater than<br />
the maximum range.<br />
Sensor Information The sensor can be queried for unit identification and quality control<br />
information.<br />
Serial Number<br />
The query gets the sensor serial number.<br />
Query: SYSTem:INFormation:SNUMber?<br />
Reply: <br />
Calibration Date<br />
The query gets the calibration date.<br />
Query: SYSTem:INFormation:CDATe?<br />
Reply: <br />
Manufacturing Date<br />
The query gets the manufacturing date.<br />
Query: SYSTem:INFormation:MDATe?<br />
Reply: