PowerMax-<strong>USB</strong>/<strong>RS</strong> User Manual<br />

Current Wavelength Command<br />

Sets the current wavelength, which is committed to persistent<br />

storage when it is changed. If the requested wavelength is greater<br />

than the upper wavelength limit, the requested wavelength is set to<br />

the upper wavelength limit. Likewise, if the requested wavelength is<br />

less than the lower wavelength limit, the requested wavelength is set<br />

to the lower wavelength limit.<br />

Command: wv <br />

Reply: <br />

Current Power Reading Query<br />

Queries the current power reading.<br />

Query: pw?<br />

Reply: <br />

Current Beam Position Query<br />

Queries the current beam position. If the sensor is an optical sensor,<br />

the returned position X and Y values are 0.<br />

Query: pos<br />

Reply (thermopile quad sensor): ,<br />

Reply (thermopile mono or optical sensor): 0,0<br />

Note that X and Y positions are zero if the sensor is a thermopile<br />

mono or if the power is less than 10% of the minimum range (see<br />

“Minimum Range Query” on page 67).<br />

Current Thermistor Counts Query<br />

Queries the current thermistor ADC value.<br />

Query: tmp<br />

Reply: <br />

Start Data Streaming Command<br />

Enables LaserPAD/SSIM mode interface data streaming. Streaming<br />

data records are transmitted at 10 Hz. Streaming data records<br />

continue to be transmitted until the Stop Data Streaming command<br />

(see below) is issued.<br />

Command: dst<br />

Reply: none<br />

68