Screen<br />

Function: Reduce wear on the screen, change<br />

the screen background appearance.<br />

Type: Menu select, Time-out 1 to 255 minutes<br />

Description: Blanks the screen when there has<br />

been no activity for a specified time period. black or white background option.<br />

Default: Screen saver is disabled, chart defaults to a white background<br />

The brightness and contrast of the screen can be controlled by a trimpot adjustment,<br />

see ”<strong>Display</strong> Trimpot Adjustments’ on page 19.<br />

The Saver function helps to reduce screen wear, highlight and toggle this option On or<br />

Off using the Enter button. Set the screen saver Timeout from 1 to 255 minutes. Highlight<br />

and select to produce a text box and enter the time-out period required.<br />

Chart paper - This feature gives the option of having a black or white chart background<br />

colour. Use the directional key to highlight and the enter button to select which<br />

chart background colour is preferred.<br />

Option Codes<br />

Function: Serial number Identification and option availability<br />

Type: 2 part 15 Character alpha/numeric.<br />

Description: 6 character serial number, factory programmed for unit identification.<br />

Plus a 9 character option code, fully upgradeable with option enhancements.<br />

Default: Individual factory set code<br />

This Option Code is a unique serial number specific to your recorder which contains<br />

information enabling certain options. It can ONLY be changed when and if you purchase<br />

further options for this recorder. The first six digits of the code depicts the serial number<br />

of the recorder. The other nine digits are the coded options available. The whole code is<br />

CRC checked, invalid entries will not be accepted.<br />

Password<br />

Function: Protects entry to the system at various levels<br />

Type: Text entry<br />

Description: Restricts access within the recorder, providing password protection at<br />

different levels.<br />

Default: Disabled<br />

Password protection restricts user entry to different levels within the recorder. Passwords<br />

will be disabled on start up and the recorder will default to a predetermined password,<br />

which is displayed. ’Eng’, is the highest level of access to all screens.<br />

The default password is ”PASS”. From here the Engineer can allocate other users and<br />

their levels of access to screens within the recorder. Each user creates their own password<br />

when they initially log on. Each user, including ‘Eng’ is responsible for remembering<br />

their own password, ‘Eng’ level cannot access the passwords for other users.<br />

NB. If the user does forget the password the user must be deleted from the recorder<br />

and start again. If ‘Eng’ user forgets their password and no access is available, contact us<br />

at SIEMENS and an override password can be issued.<br />

This password system allows traceability by the users name, not the password.<br />