SIREC DS Display Recorder 7ND4120-xxx - Siemens
SIREC DS Display Recorder 7ND4120-xxx - Siemens
SIREC DS Display Recorder 7ND4120-xxx - Siemens
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Screen<br />
Function: Reduce wear on the screen, change<br />
the screen background appearance.<br />
Type: Menu select, Time-out 1 to 255 minutes<br />
Description: Blanks the screen when there has<br />
been no activity for a specified time period. black or white background option.<br />
Default: Screen saver is disabled, chart defaults to a white background<br />
The brightness and contrast of the screen can be controlled by a trimpot adjustment,<br />
see ”<strong>Display</strong> Trimpot Adjustments’ on page 19.<br />
The Saver function helps to reduce screen wear, highlight and toggle this option On or<br />
Off using the Enter button. Set the screen saver Timeout from 1 to 255 minutes. Highlight<br />
and select to produce a text box and enter the time-out period required.<br />
Chart paper - This feature gives the option of having a black or white chart background<br />
colour. Use the directional key to highlight and the enter button to select which<br />
chart background colour is preferred.<br />
Option Codes<br />
Function: Serial number Identification and option availability<br />
Type: 2 part 15 Character alpha/numeric.<br />
Description: 6 character serial number, factory programmed for unit identification.<br />
Plus a 9 character option code, fully upgradeable with option enhancements.<br />
Default: Individual factory set code<br />
This Option Code is a unique serial number specific to your recorder which contains<br />
information enabling certain options. It can ONLY be changed when and if you purchase<br />
further options for this recorder. The first six digits of the code depicts the serial number<br />
of the recorder. The other nine digits are the coded options available. The whole code is<br />
CRC checked, invalid entries will not be accepted.<br />
Password<br />
Function: Protects entry to the system at various levels<br />
Type: Text entry<br />
Description: Restricts access within the recorder, providing password protection at<br />
different levels.<br />
Default: Disabled<br />
Password protection restricts user entry to different levels within the recorder. Passwords<br />
will be disabled on start up and the recorder will default to a predetermined password,<br />
which is displayed. ’Eng’, is the highest level of access to all screens.<br />
The default password is ”PASS”. From here the Engineer can allocate other users and<br />
their levels of access to screens within the recorder. Each user creates their own password<br />
when they initially log on. Each user, including ‘Eng’ is responsible for remembering<br />
their own password, ‘Eng’ level cannot access the passwords for other users.<br />
NB. If the user does forget the password the user must be deleted from the recorder<br />
and start again. If ‘Eng’ user forgets their password and no access is available, contact us<br />
at SIEMENS and an override password can be issued.<br />
This password system allows traceability by the users name, not the password.<br />
34 43-TV-25-20 GLO Issue 5 06/04 UK<br />