Operation and control<br />
5.10.2 Settings example normal mode<br />
With a 5% CO 2 content, a 3% oxygen content and a humidity of 96%, the appliance should<br />
heat up to 37°C. The monitoring function should respond at 38.5 °C ( Fig. 20 ).<br />
Temperature in °C<br />
50<br />
40<br />
30<br />
20<br />
10<br />
Monitoring temperature<br />
Humidity rh in %<br />
Fig. 20 Example for normal mode<br />
100<br />
80<br />
60<br />
40<br />
20<br />
CO 2 in %<br />
15<br />
10<br />
5<br />
Time Time Time<br />
1. Setting the normal operating mode:<br />
Hold SET key down for approx. 3 seconds, the current<br />
operating mode then begins to flash.<br />
Select the operating mode with the push-turn control,<br />
while the SET key is held down. After you let go of<br />
the SET key, the control is in the normal operating mode.<br />
2. Setting the temperature setpoint:<br />
Hold down the SET key and set the desired temperature<br />
setpoint of 37.0 °C with the push-turn control.<br />
Release the SET key, the appliance will briefly flash, showing<br />
the temperature setpoint. Then the current temperature<br />
appears on the display and the controller begins to move<br />
to the set temperature of 37.0 °C.<br />
► Heating up is indicated by the icon.<br />
3. Setting the monitoring temperature:<br />
Turn the push-turn control to the right until the monitoring<br />
temperature and the MIN or MAX icon flashes. Hold<br />
down the SET key and with the push-turn control, set<br />
the overtemperature limit to 38.5 °C and the undertemperature<br />
limit to 36.0 °C. Turn the push-turn control to<br />
the right until the monitoring temperature and the AUTO<br />
icons flash. Hold down the SET key and set to on with the<br />
push-turn control.<br />
The tolerance band is set in the SETUP menu (see<br />
page 45 ).<br />
25<br />
20<br />
MIN<br />
O 2 in %<br />
25<br />
20<br />
15<br />
10<br />
5<br />
AUTO<br />
Time<br />
PRINT SETUP<br />
MAX<br />
°C<br />
°C<br />
31