26.10.2012 Views

OPERATING MANUAL - Bad Request

OPERATING MANUAL - Bad Request

OPERATING MANUAL - Bad Request

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Advanced functions<br />

Example: CO 2 deviation in chamber load at 10 % should be corrected:<br />

1. Set the CO 2 balance point in the SETUP to CO2 10 and set the accompanying compensation<br />

correction value to 0.0 %:<br />

56<br />

Mo<br />

on<br />

off<br />

Tu We Th Fr Sa Su<br />

t1<br />

t2<br />

t3<br />

t4<br />

loop<br />

PRINT SETUP<br />

4<br />

3<br />

2<br />

1<br />

STERI DEFRO<br />

°C °C<br />

MIN<br />

AUTO<br />

MAX<br />

IN 1 IN 2 OUT IN 1 IN 2 OUT<br />

% rh<br />

CO2<br />

mb<br />

mb<br />

2. With a calibrated reference instrument and at a set CO 2 content of 10.0 %, an actual CO 2<br />

content of 11.5 % is measured.<br />

3. Set the compensation correction value in the SETUP for CO2 10 to 1.5 %:<br />

Mo<br />

on<br />

off<br />

Tu We Th Fr Sa Su<br />

t1<br />

t2<br />

t3<br />

t4<br />

loop<br />

PRINT SETUP<br />

4<br />

3<br />

2<br />

1<br />

STERI DEFRO<br />

°C °C<br />

MIN<br />

AUTO<br />

MAX<br />

IN 1 IN 2 OUT IN 1 IN 2 OUT<br />

% rh<br />

CO2<br />

mb<br />

mb<br />

4. The reference instrument should display 10 % after the calibration procedure.<br />

With CO2 5 and CO2 15, further calibrations can be programmed at 5 % and 15 % (15 % only<br />

for models with CO 2 or premium modules).<br />

If all compensation correction values are set to 0.0 %, the factory CO 2 calibration settings<br />

are restored.<br />

7.6.4 Oxygen calibration<br />

(only for models with O2 module)<br />

In the O calibration, the oxygen proportion in the incubator at normal atmosphere is deter-<br />

2<br />

mined.<br />

1. Switch off the O module, if active. To do this, turn the<br />

2<br />

push-turn control to the left until the O display flashes.<br />

2<br />

Hold down the set-key and turn the push-turn control<br />

Mo Tu<br />

on<br />

off<br />

We Th Fr Sa Su<br />

until O off is displayed. Release the set-key.<br />

2<br />

2. Set CO setpoint value to 0% (see page 32 ).<br />

2<br />

3. Open the oven door and inner glass door, in case of<br />

four-part door, remove glass panes. Keep the doors<br />

open for at least one minute so that a normal atmosphere<br />

can form in the chamber. Close the doors again.<br />

o off 2<br />

4. In the setup, select the item O calibration with the<br />

2<br />

push-turn control (cal 02 off). Hold down the set key<br />

and set to cal 02 on with the push-turn control. Release<br />

the set key and leave the setup via EXIT.<br />

Now, the oxygen content in the incubator is measured<br />

for about one minute. In the display, cal o2 active can be<br />

seen. Do not switch off the oven while this is taking<br />

place.<br />

5. When the calibration is finished, the oxygen content<br />

determined is shown in the display:<br />

off<br />

CAL o on 2<br />

CAL o 2 active<br />

o 2 =20.7%

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!