CMG-CD24<br />

The Broadband Noise calibration signal consists of a constant stream<br />

of white noise, which lasts for the specified number of minutes. The<br />

Period and Frequency are ignored.<br />

3.2.4 Mass control<br />

If supported by the instrument, the CMG-CD24 can initiate locking,<br />

unlocking, or centring of the mass. If the instrument does not support<br />

locking, unlocking and centring, instructing the CD24 to “centre”<br />

switches the instrument into a one-second response mode, which<br />

allows you to monitor the mass positions more easily.<br />

You can issue this instruction from the Mass Control tab:<br />

To switch into one-second response mode, click Centre.<br />

3.2.5 Data flow<br />

The CD24 operates in one of several transmission modes. These modes<br />

relate to how the unit uses its Flash memory:<br />

• as a simple data store, from which you can request data<br />

(FILING, DUAL and DUPLICATE modes);<br />

• as a buffer holding unacknowledged blocks, which are<br />

transmitted in preference to real-time data (FIFO mode); or<br />

• as a buffer holding unacknowledged blocks, which are<br />

transmitted whenever the channel is free but no real-time data<br />

blocks are ready (ADAPTIVE mode).<br />

• not at all (DIRECT mode).<br />

40 Issue C