CMG-CD24<br />

should appear in Scream!'s main window containing the<br />

returned calibration signal.<br />

3. Open a Waveview window on the calibration signal and the<br />

returned streams by selecting them and double-clicking. The<br />

streams should display the calibration signal combined with the<br />

sensors' own measurements. If you cannot see the calibration<br />

signal, zoom into the Waveview using the scaling icons at the<br />

top left of the window or the cursor keys.<br />

4. Drag the calibration stream Cn across the Waveview window, so<br />

that it is at the top.<br />

5. If the returning signal is saturated, retry using a calibration<br />

signal with lower amplitude, until the entire curve is visible in<br />

the Waveview window.<br />

6. If you need to scale one, but not another, of the traces, rightclick<br />

on the trace and select Scale.... You can then type in a<br />

suitable scale factor for that trace.<br />

7. Pause the Waveview window by clicking on the icon.<br />

8. Hold down SHIFT and drag across the window to select the<br />

calibration signal and the returning component(s). Release the<br />

mouse button, keeping SHIFT held down. A menu will pop up.<br />

If it doesn't it is likely because you have not selected at least two<br />

signals (calibration and the returning components)<br />

9. Choose Broadband Noise Calibration.<br />

50 Issue C