Operation<br />
2.4.1 Areas of Application<br />
Warning<br />
Caution<br />
Before pumping oxygen (or other highly<br />
reactive gases) at concentrations exceeding<br />
the concentration in the atmosphere<br />
(> 21 % for oxygen) it will be necessary to<br />
use a special pump. Such a pump will have<br />
to be modified and de-greased, and an inert<br />
special lubricant (like PFPE) must be used.<br />
The pump is not suitable for pumping of:<br />
– ignitable and explosive gases or vapours<br />
– oxidants<br />
– pyrophorous gases.<br />
The pumps are not suitable for pumping of<br />
liquids or very dusty media. Suitable protective<br />
devices must be installed.<br />
Our technical sales department is available for further<br />
advice in these matters.<br />
2.4.2 Notes for Pumps with<br />
EExe II T3 Motors<br />
There are no reservations as to operation of the vacuum<br />
pump in explosion hazard areas of outer ZONE 1 for<br />
substances of temperature classes T1 to T3 and explosion<br />
group IIA and IIB when observing the conditions<br />
given in the following:<br />
Warning<br />
• The electric motor of the vacuum pump<br />
must be connected through a suitable<br />
motor protection switch to the mains supply.<br />
If the vacuum pump is supplied ready<br />
for connection, the motor protection<br />
switch needs to be installed by the customer<br />
on the vacuum pump and within a<br />
pressure-tight encapsulation.<br />
• The electric motor must be installed in<br />
accordance with the accepted rules which<br />
apply to electrical systems in explosion<br />
hazard areas. If required, detailed information<br />
on this is available from the manufacturer.<br />
• In case of a failure relating to the vacuum,<br />
the vacuum pump must be switched off<br />
automatically by interrupting the power<br />
supply.<br />
• Due to the measures introduced, the<br />
vacuum pump is not suited for pumping<br />
explosive atmospheres (inner explosion<br />
protection). If planning such an application<br />
please consult us first.<br />