<strong>Quattro</strong> II<br />

<strong>User's</strong> <strong>Guide</strong><br />

The status of the system is indicated at all times by the front panel lamps:<br />

VENT<br />

PUMP<br />

Green Red Green Red Green<br />

OPERATE<br />

Red<br />

Pump Rough<br />

Flash<br />

Pump High<br />

(Standby)<br />

Operate<br />

Transient Pressure<br />

Trip (Operate)<br />

Transient Pressure<br />

Trip (Standby)<br />

RF Trip<br />

Turbo<br />

Wind Down<br />

Flash<br />

Turbo Shutdown<br />

Venting<br />

Flash<br />

Vented<br />

Protection<br />

Transient Pressure Trip<br />

The transient trip is designed to protect the instrument from potentially damaging<br />

pressure surges and operates routinely when solvent elutes from the GC column.<br />

Should the vacuum gauge(s) detect a pressure surge above the preset trip level<br />

(normally set at 10 -4 mbar by software) the following events occur:<br />

Routine Procedures<br />