<strong>Quattro</strong> II<br />

<strong>User's</strong> <strong>Guide</strong><br />

Carefully slide the RF lens assembly along the guide rods until the inner end is<br />

felt to engage in the analyser entrance orifice.<br />

RF Lens<br />

Assembly<br />

API Adaptor<br />

Flange<br />

Release the quick release clamp and remove the blanking flange. Retain the O<br />

ring and clamp.<br />

Insert the API source housing into the analyser housing and secure with the two<br />

black thumb nuts.<br />

Using the O ring and quick release clamp, connect the pumping tube to the port<br />

on the side of the source.<br />

Check that the manual vent valve is closed.<br />

Routine Procedures<br />