Technology and maintenance – Chapter 6<br />
Electromagnetic immunity I<br />
The <strong>MediBalance</strong> <strong>Pro</strong> is intended for use in the electromagnetic environment specifi ed below. The customer<br />
or the user of the <strong>MediBalance</strong> <strong>Pro</strong> should assure that it is used in such environment.<br />
Immunity test IEC 60601 test<br />
level<br />
Compliance level Electromagnetic environment -<br />
guidance<br />
Electrostatic discharge<br />
(ESD) IEC 61000-4-2<br />
Electrical fast transient<br />
/ burst IEC 61000-4-4<br />
6 kV contact<br />
8 kV air<br />
2 kV for power<br />
supply lines<br />
1 kV for input /<br />
output lines<br />
Surge IEC 61000-4-5 1 kV common<br />
mode<br />
2 kV differential<br />
mode<br />
Voltage dips, short interruptions<br />
and voltage<br />
variations on power<br />
supply input lines IEC<br />
61000-4-11<br />
Power frequency<br />
(50/60 Hz) magnetic<br />
fi eld IEC 61000-4-8<br />
< 5 % U T<br />
(> 95% dip of U T<br />
)<br />
for 1/2 cycle<br />
40 % U T<br />
(60 % dip of U T<br />
)<br />
for 5 cycle<br />
70 % U T<br />
(30 % dip of U T<br />
)<br />
for 25 cycle<br />
95 % dip of U T<br />
)<br />
for 5 s<br />
6 kV contact<br />
8 kV air<br />
not applicable<br />
not applicable<br />
not applicable<br />
Floors should be wood, concrete or<br />
ceramic tile. If fl oors are covered with<br />
synthetic material, the relative humidity<br />
shall be al least 30 %.<br />
Mains power quality should be similar<br />
to that of a typical commercial or hospital<br />
environment.<br />
Mains power quality should be similar<br />
to that of a typical commercial or hospital<br />
environment.<br />
Mains power quality should be similar<br />
to that of a typical commercial or<br />
hospital environment. When the user<br />
of the <strong>MediBalance</strong> <strong>Pro</strong> continued<br />
function also calls in the envent of<br />
disruption of supply, it is recommended<br />
the <strong>MediBalance</strong> <strong>Pro</strong> form an<br />
uninterruptible power supply.<br />
3 A/m 3 A/m Power frequency magnetic fi elds<br />
should be at leves characteristic of a<br />
typical location in a typical emmercial<br />
or hospital environment.<br />
Note: U T<br />
is the ac mains voltage prior to application of the test level.<br />
36