2.4.3 Meter Configuration<br />
The <strong>Model</strong> <strong>1835</strong>-C provides a method to save the configuration of the meter<br />
and to recall that configuration for later use even if the meter has been turned<br />
off. This is accomplished through configuration buffers maintained in nonvolatile<br />
memory. Configuration buffers are numbered 0 to 9 with buffer 0<br />
being a DEFAULT buffer which can only be recalled but not saved to. The<br />
reset state of all the buffers except the default buffer is empty. Empty buffers<br />
cannot be recalled.<br />
A recalled configuration becomes the current configuration of the meter. Any<br />
changes to the current configuration must be saved via SAVE CONFIG or they<br />
will be lost when a new configuration is recalled.<br />
The meter will not recall a configuration that is not compatible with the<br />
detector calibration module currently plugged into the meter. Configurations<br />
using the same model of detector are compatible while configurations using<br />
different models of detectors are incompatible.<br />
The default configuration of the meter depends upon the detector family. The<br />
list of configuration parameters stored in a configuration buffer as well as<br />
their default values versus detector family are listed in Table 9 below:<br />
Table 9. Configuration Parameters and Default Conditions<br />
Parameter Detector Family Default Condition<br />
MODE Low-Power, High-Power DC CONT<br />
Energy<br />
CONT PULSE<br />
UNITS Low-Power, High-Power W<br />
Energy<br />
J<br />
λ, Lambda Lowest available<br />
PRESENT RESP<br />
PRESENT RESP<br />
USR CAL<br />
OFF<br />
ATTN<br />
OFF<br />
FILTER<br />
OFF<br />
DC SAMPLE PREC Low-Power, High-Power 20,000 CNT<br />
DC SAMPLE FREQ<br />
25 Hz<br />
AUTO<br />
ON<br />
RANGE 0<br />
ZERO 0.000<br />
Zero Value 0<br />
Reference Value 0.001<br />
STATS Buffer Size 10<br />
DET SWITCH POS Energy I, (Intermediate)<br />
DATA STORE<br />
OFF<br />
DATA STORE BUFFER<br />
SLIDE<br />
D_BUF_SIZE 100<br />
15