Configuration<br />
Output 2<br />
Temperature unit and probe. Output current.<br />
Menu group<br />
Output 2 o2.<br />
Code Display<br />
Select menu item<br />
Select °C/°F<br />
Select temperature probe<br />
Select 0-20 / 4-20 mA<br />
Enter current beginning<br />
Enter current end<br />
Set output filter<br />
22 mA for temp error<br />
Hold mode<br />
Code Display Action Choices<br />
o2. Specify temperature unit<br />
Select with key<br />
Proceed with enter<br />
Select temperature probe<br />
Select with key<br />
Proceed with enter<br />
Set output current range<br />
Select with key<br />
Proceed with enter<br />
Current beginning:<br />
Enter lower end of scale.<br />
Select with key, edit number<br />
with key, proceed with enter<br />
Current end:<br />
Enter upper end of scale.<br />
Select with key, edit number<br />
with key, proceed with enter<br />
°C<br />
(°F)<br />
22NTC<br />
(30NTC)<br />
4 - 20 mA<br />
0 - 20 mA<br />
000.0 °C<br />
100.0 °C<br />
Process temperature: Current beginning and current end<br />
Example 1: Range 0 to 100 °C<br />
[°C]<br />
100 Process temperature<br />
Example 2: Range 50 to 70 °C.<br />
Advantage: Higher resolution in<br />
range of interest<br />
[°C]<br />
70<br />
Process temperature<br />
End:<br />
Press conf, then enter<br />
38 O 2 4100 e<br />
0<br />
4<br />
Output current<br />
20<br />
[mA]<br />
50<br />
4<br />
Output current<br />
20 [mA]<br />
39