Configuration<br />
Output 2<br />
Temperature error. Output current during HOLD.<br />
Menu group<br />
Output 2 o2.<br />
Code Display<br />
Select menu item<br />
Select °C/°F<br />
Select temperature probe<br />
Select 0-20 / 4-20 mA<br />
Enter current beginning<br />
Enter current end<br />
Set output filter<br />
22 mA for temp error<br />
Hold mode<br />
Code Display Action Choices<br />
o2. 22 mA signal for error message<br />
Select with key<br />
Proceed with enter<br />
Output signal during HOLD<br />
LAST: During HOLD the last measured<br />
value is maintained at the<br />
output<br />
FIX: During HOLD a value (to be<br />
entered) is maintained at the output<br />
Select with key<br />
Proceed with enter<br />
Only with FIX selected:<br />
Enter current which is to flow at<br />
the output during HOLD<br />
Select position with key and<br />
edit number with key<br />
Proceed with enter<br />
OFF<br />
(ON)<br />
LAST<br />
(FIX)<br />
21.0 mA<br />
(00.0 ...<br />
21.0 mA<br />
Output signal for HOLD:<br />
Output current<br />
[mA]<br />
21<br />
Output signal HOLD<br />
Setting FIX = 21.0 mA<br />
Output signal HOLD<br />
Setting LAST<br />
End:<br />
Press conf, then enter<br />
42 O 2 4100 e<br />
4<br />
HOLD active<br />
HOLD active<br />
43