Error messages (error codes)<br />
Errors<br />
ERR 01<br />
Display Problem<br />
Possible causes<br />
Measured<br />
value<br />
flashes<br />
SAT range<br />
Sensor defective<br />
Wrong sensor connected<br />
Measurement range exceeded<br />
Alarm contact<br />
Red LED<br />
Out 1 (22 mA)<br />
Out 2 (22 mA)<br />
x x x<br />
Errors<br />
ERR 03<br />
ERR 11<br />
Symbol<br />
(flashing)<br />
Problem<br />
Possible causes<br />
Temperature probe<br />
Open or short circuit<br />
Temperature range exceeded<br />
Current output 1<br />
Current below 0 (3.8) mA<br />
Alarm contact<br />
x<br />
x<br />
Red LED<br />
x<br />
x<br />
Out 1 (22 mA)<br />
x<br />
x<br />
Out 2 (22 mA)<br />
x<br />
ERR 02<br />
Measured<br />
value<br />
flashes<br />
Conc range<br />
Sensor defective<br />
Wrong sensor connected<br />
Measurement range exceeded<br />
x<br />
x<br />
x<br />
ERR 12<br />
ERR 13<br />
Current output 1<br />
Current above 20.5 mA<br />
Current output 1<br />
Current span too small / too large<br />
x<br />
x<br />
x<br />
x<br />
x<br />
x<br />
ERR 21<br />
Current output 2<br />
Current below 0 (3.8) mA<br />
x<br />
x<br />
x<br />
ERR 98<br />
“ConF”<br />
flashes<br />
System error<br />
Configuration or calibration data<br />
defective; completely reconfigure<br />
and recalibrate the device<br />
Memory error in device program<br />
(PROM defective)<br />
x<br />
x<br />
x<br />
x<br />
ERR 22<br />
ERR 23<br />
ERR 41<br />
Current output 2<br />
Current above 20.5 mA<br />
Current output 2<br />
Current span too small / too large<br />
Rinsing probe<br />
Communication error<br />
x<br />
x<br />
x<br />
x<br />
x<br />
x<br />
x<br />
x<br />
x<br />
x<br />
ERR 99<br />
“FAIL”<br />
flashes<br />
Factory settings<br />
EEPROM or RAM defective<br />
This error message only occurs in<br />
the case of a complete defect. The<br />
Transmitter must be repaired and<br />
recalibrated at the factory.<br />
x<br />
x<br />
x<br />
x<br />
ERR 33<br />
Sensocheck<br />
Sensor: Connecting cable defective<br />
• Zero error, Sensoface active, see Pg 82<br />
• Slope error, Sensoface active, see Pg 82<br />
• Response time exceeded, Sensoface active, s. Pg 82<br />
x<br />
x<br />
x<br />
• Calibration interval expired, Sensoface active, s. Pg 82<br />
78<br />