04.11.2012 Views

OPERATING MANUAL - Hettich AG, CH

OPERATING MANUAL - Hettich AG, CH

OPERATING MANUAL - Hettich AG, CH

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Design and description<br />

Important: The relative air humidity can only be set in the temperature range of 10 °C to 95 °C.<br />

The CTC climatic test chamber has two freshwater tanks with automatic switchover (see page<br />

22 ). The currently active tank is indicated in the control display by “TANK1” or “TANK2” (see<br />

page 28 ).<br />

2.3 Working range<br />

Caution:<br />

If in operation at the upper level or outside the working range for long periods, puddles<br />

of water may form inside the chamber and water may pass the door seal.<br />

Humidity rh%<br />

12<br />

100<br />

90<br />

80<br />

70<br />

60<br />

50<br />

40<br />

30<br />

20<br />

10<br />

0<br />

0 10 20 30 40 50 60 70 80 90 100<br />

Temperature in ˚C<br />

Fig. 5 Control range of temperature and<br />

relative humidity<br />

°C<br />

Temperature in<br />

200<br />

150<br />

100<br />

50<br />

0<br />

-50<br />

0 10 20 30 40 50 60 70 80 90 100<br />

min<br />

180 to -40 degrees in 80 minutes<br />

Fig. 7 Cooling / heating up speed<br />

Heat input [W]<br />

Temperature in ˚C<br />

Fig. 6 Heat compensation T ambient = 25 °C<br />

Temperature in ˚C<br />

200<br />

150<br />

100<br />

50<br />

0<br />

-50<br />

0 10 20 30 40 50 60 70 80 90 100<br />

min<br />

-40 to 180 degrees in 20 minutes