Operation and control<br />

Direct setting of the temperature setpoint:<br />

If the controller is in standby mode or the week time switch is in the ON phase, the temperature<br />

setpoint can be directly accessed by briefly pressing the SET key. By turning the control to<br />

the right, you are returned to temperature monitoring and humidity. By turning to the left you<br />

come back to the settings for the individual time blocks.<br />

5.8.4 Example of settings Week time switch<br />

The test chamber should be switched on at 07:30 from Mo–Fr (working day group) and<br />

switched off at 18:00, and in addition it should operate on Saturdays from 10:00–14:00 ( Fig.<br />

23 ).<br />

34<br />

Mo<br />

Tu<br />

We<br />

Thu<br />

Fig. 23 Operation with week time switch (example)<br />

1. Setting the week time switch operating mode<br />

Hold the SET key down for approx. 3 seconds; the current<br />

operating mode then begins to flash. Select the week<br />

time switch operating mode with the push-turn control,<br />

while the SET key is held down.<br />

Release the SET key; the control is now in the week time<br />

switch operating mode.<br />

2. Switch on Mo-Fr at 7:30<br />

Turning the push-turn control to the left, select “Mo-Fr<br />

on” (group working days).<br />

Hold down the SET key and set the desired switch-on<br />

time to 7:30 with the push-turn control.<br />

Fr<br />

Sa<br />

Su<br />

Mo Tu We Th Fr<br />

on<br />

off<br />

PRINT SETUP<br />

Sa Su<br />