Advanced functions<br />

The ASF alarm goes off automatically as soon as 50 % of the set tolerance band of the setpoint<br />

(in the example 50 °C ± 1 °C) is reached again (section A).<br />

If the temperature setpoint is altered, the ASF is automatically disabled temporarily (see in<br />

the example: The setpoint is changed from 50 °C to 25 °C, section D), until it has reached the<br />

tolerance range of the new temperature setpoint (section E).<br />

50<br />

°C<br />

50°C<br />

25°C<br />

52°C<br />

48°C<br />

ASF active<br />

A B C D E<br />

ASF alarm<br />

52°C<br />

48°C<br />

27°C<br />

23°C<br />

AUTO ASF active AUTO ASF active AUTO<br />

Fig. 28 Schematic diagram of how the ASF temperature monitoring functions<br />

Switching on the automatic temperature monitor:<br />

Select the AUTO icon with the push-turn control.<br />

Hold down the SET key and set to on with the push-turn control.<br />

Switching off the automatic temperature monitor:<br />

Select the AUTO icon with the push-turn control.<br />

Hold down the SET key and set to off with the push-turn control.<br />

The tolerance band for the ASF can be set in the SETUP in the range 2 ... 20 °C (see Chapter<br />

7.2 Basic appliance settings (Setup) on page 46 ).<br />

MIN<br />

MIN<br />

AUTO<br />

AUTO<br />

t<br />

MAX<br />

MAX<br />

°C<br />

°C