Advanced functions<br />

52<br />

CAL 1<br />

+0,5 °C<br />

-40 °C<br />

CAL 2<br />

-0,4 °C<br />

30°C<br />

Fig. 29 Temperature calibration (example)<br />

CAL 3<br />

+0,8 °C<br />

Calibration at factory<br />

150 °C<br />

Setting:<br />

1. Set the desired compensation temperature in the SETUP (see Chapter 7.2 Basic appliance<br />

settings (Setup) on page 46 ) and set the corresponding compensation correction value to<br />

0.0 °C.<br />

2. With a reference instrument, measure the deviation in the stationary state in the selected<br />

compensation temperature.<br />

3. Setting the compensation correction values in the SETUP. If the measured reference temperature<br />

is too low, the compensation correction value must be set with a negative sign.<br />

4. Perform a control measurement with the reference instrument.<br />

5. Repeat the procedure for the other two balance points if necessary.<br />

Example: Temperature deviation at 30 °C should be corrected.<br />

1. Set compensation temperature CAL.2 in the SETUP to 30.0 °C and set the accompanying<br />

compensation correction value to 0.0 °C:<br />

h<br />

°C °C<br />

2. At a set setpoint temperature of 30 °C in normal operation, an actual temperature of 29.6<br />

°C is measured with a calibrated reference instrument.<br />

3. Set the compensation correction value for CAL.2 in the SETUP to – 0.4 °C:<br />

h<br />

°C °C