Advanced functions<br />
52<br />
CAL 1<br />
+0,5 °C<br />
-40 °C<br />
CAL 2<br />
-0,4 °C<br />
30°C<br />
Fig. 29 Temperature calibration (example)<br />
CAL 3<br />
+0,8 °C<br />
Calibration at factory<br />
150 °C<br />
Setting:<br />
1. Set the desired compensation temperature in the SETUP (see Chapter 7.2 Basic appliance<br />
settings (Setup) on page 46 ) and set the corresponding compensation correction value to<br />
0.0 °C.<br />
2. With a reference instrument, measure the deviation in the stationary state in the selected<br />
compensation temperature.<br />
3. Setting the compensation correction values in the SETUP. If the measured reference temperature<br />
is too low, the compensation correction value must be set with a negative sign.<br />
4. Perform a control measurement with the reference instrument.<br />
5. Repeat the procedure for the other two balance points if necessary.<br />
Example: Temperature deviation at 30 °C should be corrected.<br />
1. Set compensation temperature CAL.2 in the SETUP to 30.0 °C and set the accompanying<br />
compensation correction value to 0.0 °C:<br />
h<br />
°C °C<br />
2. At a set setpoint temperature of 30 °C in normal operation, an actual temperature of 29.6<br />
°C is measured with a calibrated reference instrument.<br />
3. Set the compensation correction value for CAL.2 in the SETUP to – 0.4 °C:<br />
h<br />
°C °C