Figure 4-2<br />

.<br />

For <strong>BD</strong> FACS 7-color setup beads, locate this information on the foil<br />

reagent packet, the target values sticker, or the side of the reagent box.<br />

Do not use the numbers on the setup beads tube.<br />

4 In the Setup Lot Information dialog, enter the target values for the bead lot.<br />

To change a target value, select the current value in the Target Value field<br />

and enter the new value. Repeat until you have edited all target values.<br />

Target values determine where the software places the setup beads during<br />

detector and spectral overlap adjustments. They must be edited for every<br />

new bead lot. Target values are provided with each box of beads. See<br />

Figure 4-3 on page 47 for an example setup beads label.<br />

46 <strong>BD</strong> <strong>FACSCanto</strong> <strong>Clinical</strong> <strong>Software</strong> <strong>Reference</strong> <strong>Manual</strong>