<strong>MI</strong> <strong>019</strong>-<strong>100</strong> – December 2003<br />
2. Installation<br />
♦<br />
♦<br />
♦<br />
provide a minimum of 25 mm (1 in) of axial length of surface contact. “Rest type”<br />
supports should not be used with sanitary end connections.<br />
If your application has excessive vibration that could be transferred to the flowtube<br />
sensor, install isolation supports such as those manufactured by Stauff or Behringer.<br />
Position valves outside the pipe supports described above.<br />
For severe vibration conditions, install two supports on each side of flowtube as<br />
shown in Figure 4.<br />
FLANGED END CONNECTIONS<br />
FLOWTUBE<br />