3. Start Up<br />
Procedure Using the HART Communicator<br />
The HART Communicator Quick Start Menu is shown in Figure 20.<br />
1. Go to 2 Online.<br />
2. Go to 2 Quick Start.<br />
3. Go to 1 <strong>Flow</strong> Constants and enter your flow constants.<br />
4. Go to 2 Density Constants and enter your density constants.<br />
5. Apply flow to your flowmeter for 5 to 10 minutes.<br />
6. Create zero flow by closing block valves to ensure no fluid movement.<br />
7. Go to 3 <strong>Flow</strong> Zero and zero your flowmeter.<br />
8. Go to 4 URV and enter your upper range value.<br />
9. Go to 5 LRV and enter your lower range value.<br />
2 Online<br />
1 Measure<br />
2 Quick Start 1 <strong>Flow</strong> Constants<br />
2 Density Constants<br />
3 <strong>Flow</strong> Zero<br />
4 URV<br />
5 LRV<br />
Figure 20. HART Communicator Quick Start Menu<br />
28