3. Start Up<br />

Procedure Using the HART Communicator<br />

The HART Communicator Quick Start Menu is shown in Figure 20.<br />

1. Go to 2 Online.<br />

2. Go to 2 Quick Start.<br />

3. Go to 1 <strong>Flow</strong> Constants and enter your flow constants.<br />

4. Go to 2 Density Constants and enter your density constants.<br />

5. Apply flow to your flowmeter for 5 to 10 minutes.<br />

6. Create zero flow by closing block valves to ensure no fluid movement.<br />

7. Go to 3 <strong>Flow</strong> Zero and zero your flowmeter.<br />

8. Go to 4 URV and enter your upper range value.<br />

9. Go to 5 LRV and enter your lower range value.<br />

2 Online<br />

1 Measure<br />

2 Quick Start 1 <strong>Flow</strong> Constants<br />

2 Density Constants<br />

3 <strong>Flow</strong> Zero<br />

4 URV<br />

5 LRV<br />

Figure 20. HART Communicator Quick Start Menu<br />