<strong>Ready</strong> <strong>for</strong> <strong>Recycling</strong> <strong>Instructions</strong><br />

VITROS ® 5600 Integrated System<br />

Pub. No.: J39845 EN<br />

Issued: 2011-02-18<br />

Purpose<br />

To prepare the VITROS ® 5600 Integrated System <strong>for</strong> safe end-of-life transport and recycling. This is<br />

in compliance with Directive 2002/96/EC of the European Union (EU) on Waste Electrical and<br />

Electronic Equipment (WEEE).<br />

Note: If it is not possible to follow this decontamination procedure, please contact Ortho Clinical<br />

Diagnostics.<br />

Important Considerations Be<strong>for</strong>e Starting<br />

• Assume that all equipment is contaminated with potentially infectious biological material<br />

and observe the following standards:<br />

— The “Universal Precautions” recommended by the World Health Organization<br />

(WHO) and the Centers <strong>for</strong> Disease Control (CDC).<br />

— Directive 2000/54/EC of the EU “on the protection of workers from risks related<br />

to exposure to biological agents at work”.<br />

— Standard 29CFR1910.1030, “Bloodborne Pathogens” of the Occupational<br />

Safety and Health Administration (OSHA).<br />

Obtain copies of these standards at the website <strong>for</strong> each organization.<br />

• Use the following personal protective equipment <strong>for</strong> this procedure:<br />

— disposable safety gloves<br />

— safety glasses<br />

— lab coat<br />

• Follow the manufacturer’s safety precautions <strong>for</strong> handling isopropyl alcohol.<br />

• Place all discarded system parts and cleaning materials in a biohazard disposal container.<br />

• Remove any dried test specimens or reagents when preparing the system <strong>for</strong> transport.


Assemble Cleaning Supplies<br />

Note: This in<strong>for</strong>mation is from the System Cleaning procedures.<br />

Depending upon the procedure, you will need the following materials:<br />

• disposable cloth or paper towels<br />

• warm, soapy water<br />

• Maintenance pack (OCD catalog number 1831312)<br />

• Universal wash reagent (OCD catalog number 8389793)<br />

• FL-80 cleaning solution (OCD catalog number 1877158)<br />

• Three signal reagent (SR) packs<br />

• 70% isopropyl alcohol-in-water solution<br />

• Deionized or distilled water<br />

Note: Use soap supplies in the laboratory specific <strong>for</strong> laboratory use.<br />

Cleaning Procedure<br />

IMPORTANT:<br />

REFER TO V-DOCS TO LOCATE THE DAILY, WEEKLY, MONTHLY AND AS-<br />

REQUIRED PERIODIC MAINTENANCE PROCEDURES AND FOLLOW THE<br />

CLEANING INSTRUCTIONS FOR EACH OF THE ITEMS LISTED BELOW.<br />

1. Follow the instructions displayed on the “Periodic Maintenance - Daily” and “Periodic<br />

Maintenance - Weekly” screens to clean the system.<br />

2. Follow the instructions displayed on the “Periodic Maintenance - Monthly” and “Periodic<br />

Maintenance As-Required” screens to per<strong>for</strong>m System Cleaning.<br />

3. Follow the instructions displayed on the screens to per<strong>for</strong>m the following procedures to<br />

continue cleaning the system:<br />

• Clean PM discard chute<br />

• Clean and then discard the PM evaporation caps<br />

• Clean MicroSensor cover and Ring Area<br />

• Inspect/Clean Reagent Supply Top Cover<br />

• Inspect/Clean Supply 3 Pack Opener<br />

• Inspect/Clean and then discard the reagent Cooler Filter<br />

• Inspect/Clean and then discard the Master Computer Filter<br />

• Clean System Cabinetry<br />

• Inspect/Clean Cuvette Supply<br />

• Inspect/Clean MicroTip Supply<br />

• Clean cuvetip ring area<br />

• Clean primary tip sealer<br />

• Clean secondary tip sealer<br />

2 <strong>Ready</strong> <strong>for</strong> <strong>Recycling</strong> <strong>Instructions</strong> J39845 EN<br />

VITROS ® 5600 Integrated System 2011-02-18


• Clean cuvette incubator<br />

• Clean cuvette transport arm<br />

• Inspect/Clean the Microwell Incubator area <strong>for</strong> debris and clean the incubator surfaces,<br />

if necessary.<br />

4. Clean the system cabinet with warm, soapy water. Rinse with clean water and wipe dry.<br />

5. Clean the system cabinet using 70% isopropyl alcohol and allow to air dry.<br />

6. Open the top covers and clean the interior surfaces using a cloth and 70% isopropyl alcohol.<br />

Remove and Discard the Disposable Items<br />

Final Steps<br />

Revision History<br />

7. Remove and dispose of tips from the ERF, IWF, and metering.<br />

8. Remove any debris from the side incubator area and clean the incubator surfaces.<br />

• the desiccant and humidification packs<br />

• the ERF and IWF reservoirs<br />

• all proboscises<br />

• all filters (Master Computer, System, and Cooler)<br />

• all slide cartridges<br />

• all reagent packs<br />

• any tips that might have dropped into the system<br />

• the sample trays<br />

9. If present, remove and dispose of the waste containers.<br />

10. Remove and dispose of the liquid wash and waste bottles.<br />

11. Remove the paper from the printer.<br />

12. Disconnect the power from the system.<br />

After cleaning, set the device aside <strong>for</strong> three weeks. At the end of three weeks, the device is ready<br />

<strong>for</strong> recycling and may enter the WEEE stream. Contact your local recycling or collection system <strong>for</strong><br />

further in<strong>for</strong>mation.<br />

Date of Revision Version Description of Changes<br />

2011-02-18 1.0 First release of document<br />

J39845 EN <strong>Ready</strong> <strong>for</strong> <strong>Recycling</strong> <strong>Instructions</strong> 3<br />

2011-02-18 VITROS ® 5600 Integrated System


Ortho Clinical Diagnostics<br />

100 Indigo Creek Drive<br />

Rochester, NY 14626<br />

Pub. No.: J39845 EN<br />

All trademarks are registered to<br />

Ortho-Clinical Diagnostics, Inc.<br />

© Ortho-Clinical Diagnostics, Inc. 2011

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!