<strong>Site</strong> <strong>Specifications</strong><strong>for</strong> <strong>the</strong>VITROS ® 5,1 <strong>FS</strong> <strong>Chemistry</strong> <strong>System</strong><strong>and</strong> VITROS ® 4600 <strong>Chemistry</strong> <strong>System</strong>Publication No. J398972014-03-01Supersedes J398972012-11-21


PLEASE NOTEThe in<strong>for</strong>mation contained herein is based on <strong>the</strong> experience <strong>and</strong> knowledge relating to <strong>the</strong> subject mattergained by Ortho-Clinical Diagnostics, Inc. (OCD) prior to publication.No patent license is granted by this in<strong>for</strong>mation.OCD reserves <strong>the</strong> right to change this in<strong>for</strong>mation without notice, <strong>and</strong> makes no warranty, expressed orimplied, with respect to this in<strong>for</strong>mation. OCD shall not be liable <strong>for</strong> any loss or damage, includingconsequential or special damages, resulting from any use of this in<strong>for</strong>mation, even if loss or damage iscaused by <strong>the</strong> negligence or o<strong>the</strong>r fault of OCD.This equipment includes parts <strong>and</strong> assemblies sensitive to damage from electrostaticdischarge. Use caution to prevent damage during all service procedures.Table of ContentsDescriptionPageElectrical Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Electrical <strong>Specifications</strong> <strong>for</strong> <strong>the</strong> 5,1 <strong>and</strong> 4600 <strong>System</strong>s . . . . . . . . . . . . . . . . . . . . . . . . 4Electrical <strong>Specifications</strong> <strong>for</strong> <strong>the</strong> Supplied PRINTER . . . . . . . . . . . . . . . . . . . . . . . . . . 5Electrical Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6RECEPTACLE Location <strong>Specifications</strong> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Environmental <strong>Specifications</strong> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Networking <strong>Specifications</strong> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Requirements at Customer <strong>Site</strong> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Shipping <strong>Specifications</strong> . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11<strong>Specifications</strong> <strong>for</strong> <strong>the</strong> <strong>System</strong> <strong>and</strong> <strong>the</strong> PRINTER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12<strong>Specifications</strong> <strong>for</strong> <strong>the</strong> 5,1 <strong>and</strong> 4600 HEAT REJECTION PLENUM . . . . . . . . . . . . . . 132 2014-03-01 – J39897


Section 1: Electrical Power RequirementsElectrical Power RequirementsAC Power QualityThe VITROS ® 5,1 <strong>FS</strong> <strong>Chemistry</strong> <strong>System</strong> (hereafter referred to as <strong>the</strong> 5,1) <strong>and</strong> <strong>the</strong> VITROS ® 4600 <strong>Chemistry</strong> <strong>System</strong>(hereafter referred to as <strong>the</strong> 4600) are designed to per<strong>for</strong>m satisfactorily with reasonable disturbances <strong>and</strong> voltagetolerance to <strong>the</strong> ac power. The quality <strong>and</strong> reliability of <strong>the</strong> power source is often dependent on <strong>the</strong> geographiclocation of <strong>the</strong> site. Some sites, where power quality is exceptionally poor, may require special power conditioningequipment to ensure that <strong>the</strong> tolerances are maintained within acceptable limits. Some models ofUNINTERRUPTIBLE POWER SUPPLIES (UPS) are recommended as optional equipment <strong>for</strong> use with VITROS ®<strong>System</strong>s, <strong>and</strong> may be purchased from <strong>the</strong> Distribution Center. For additional in<strong>for</strong>mation, see <strong>the</strong> UPS User Guidelocated at: http://apps.orthoclinical.com/notes/contentpages/techdoc.aspx?view=USOTHER&culture=en-usStatic ElectricityThe <strong>System</strong> should be adequately grounded to prevent <strong>the</strong> effects of static charges generated. Static problems canalso be reduced by ensuring that <strong>the</strong> relative humidity of <strong>the</strong> room is maintained above 30 percent. Special anti-staticfloor pads can be purchased <strong>and</strong> positioned where <strong>the</strong> operator is required to make contact with <strong>the</strong> instrument.Electromagnetic InterferenceSeveral types of electrical interference may be indigenous to <strong>the</strong> site location <strong>and</strong> may require special filtering toprevent malfunctions in <strong>the</strong> 5,1 or 4600 systems. If interference is suspected to be a problem, shielding may berequired, or filtering of <strong>the</strong> ac line power to <strong>the</strong> site may be necessary.Cellular phones, two-way pagers, <strong>and</strong> o<strong>the</strong>r RF transmitting devices should not be used within 1 m of <strong>the</strong> 5,1 or 4600<strong>System</strong>s.2014-03-01 – J39897 3


Section 2: Electrical <strong>Specifications</strong> <strong>for</strong> <strong>the</strong> 5,1 <strong>and</strong> 4600 <strong>System</strong>sInput Voltage* 200 to 240 VInput Frequency Limits 50 - 60 Hz 2 HzPowerOperating (Normal) 1225 WattsMaximum Current 15APOWER CORD 4 m (13 ft)PLUG TypeNorth America IEC 320 to NEMA L6-20PContinental Europe IEC 320 to CEE7 “Schuko”O<strong>the</strong>rAs required by local electrical codesCIRCUIT Type Dedicated with Isolated GroundRECEPTACLE TypeNorth America NEMA L6-20RContinental Europe CEE7 “Schuko”O<strong>the</strong>rAs required by local electrical codes* When measuring supply voltage:• Use a MULTIMETER.• Check a number of times <strong>and</strong> use <strong>the</strong> average of <strong>the</strong> values.• Wait a minimum of 5 minutes between checks.• The difference between <strong>the</strong> values should not be more than 6%.ImportantIf an optional UPS is to be installed, make certain that<strong>the</strong> correct receptacle is installed in <strong>the</strong> lab.UPS Rating Amps Receptacle Type3 KVA 20 A L6-20R**5 KVA 30 A L6-30R**** or an equivalent to meet local electrical codes208 V Check between Voltage60 Hz L1 - L2 190 - 220 V ACL1 - SAFETY GROUND 113 - 127 V ACL2 - SAFETY GROUND 113 - 127 V AC240 V Check between Voltage60 Hz L1 - L2 226 - 254 V ACL1 - SAFETY GROUND 113 - 127 V ACL2 - SAFETY GROUND 113 - 137 V ACNoteCEE7 “Schuko” plug may be supplied, ei<strong>the</strong>r straight or right angle.4 2014-03-01 – J39897


Electrical <strong>Specifications</strong> <strong>for</strong> <strong>the</strong> Supplied PRINTERSection 3: Electrical <strong>Specifications</strong> <strong>for</strong> <strong>the</strong> Supplied PRINTERImportantThe following in<strong>for</strong>mation is only applicable to PRINTERS supplied by OCD. If <strong>the</strong> PRINTER is being supplied by<strong>the</strong> customer, refer to <strong>the</strong> manufacturer’s PRINTER manual <strong>for</strong> <strong>the</strong> electrical specifications.1 RECEPTACLErequired3-wire, 120 V AC, polarized, SAFETY GROUND,maximum distance of 182.7 cm (72 in.) from eachside of <strong>the</strong> system.DedicatedCIRCUIT Type120 V Check between Voltage60 Hz N - L1 102 - 132 V AC15 amperes L1 - SAFETY GROUND 102 - 132 V ACN - SAFETY GROUND Less than 1 V ACDetach 1NEMA 5-15RSAFETYGROUNDL1Npolarized powerRECEPTACLEE002_1331ACAE002_1331AC2014-03-01 – J39897 5


Section 4: Electrical RequirementsTension200 to 240 VoltsIntensity15.0 Amps MaximumFrequency50 / 60 HzConsumption (nominal) 3.0 kVAMax Power3000 WNumber of socket-outlet 1, plus 1 <strong>for</strong> printerProtection against micro cuts Yes(Momentary power interruption)Maximum timeTolerant of missing one full cycle of 50 Hz(20 milliseconds)<strong>System</strong> tolerance to electrical 180 to 264 VAC 48-62 Hz extremevariation6 2014-03-01 – J39897


Section 5: RECEPTACLE Location <strong>Specifications</strong>RECEPTACLE Location <strong>Specifications</strong><strong>Specifications</strong> <strong>for</strong> PRINTER POWER <strong>and</strong>NETWORK CONNECTIONReferenceQ (heightfrom floor)S (printerpower)T(Network)Description30.5 cm (12 in.) minimum61.0 cm (24 in.) maximum2.9 m (8 ft. 10 in.) maximum5.3 m (17 ft. 6 in.) maximum<strong>Specifications</strong> <strong>for</strong> <strong>the</strong> <strong>System</strong> onlyReferenceQ (heightfrom floor)R (systempower)Description30.5 cm (12 in.) minimum61.0 cm (24 in.) maximum1.98 m (6 ft. 6 in.) maximumDetach 22014-03-01 – J39897 7


Section 6: Environmental <strong>Specifications</strong>Acoustic Noise Level 61 dbThermal CharacteristicsOn 4187 Btu/hrDetach 3Maximum altitude: 2438 m (8000 ft.)This product has been evaluated <strong>and</strong> is suitable only <strong>for</strong> Pollution Degree II <strong>and</strong> Insulation Category II.Additional Requirements <strong>for</strong> <strong>the</strong> VITROS ® 5,1 <strong>and</strong> VITROS ® 4600 HEAT REJECTION PLENUMIf <strong>the</strong> customer chooses to purchase <strong>and</strong> use <strong>the</strong> optional Heat Rejection Plenum, <strong>the</strong> EXHAUST SYSTEM <strong>and</strong>DUCT must be installed by <strong>the</strong> customer.<strong>Specifications</strong>Air Capacity7.1 - 9.9 cubic metres (250 - 350 cubic ft.) per minute259 - 366 metres (850 - 1200 ft.) per minuteDUCT20.5 cm (8 in.) flexible AIR DUCT <strong>and</strong> attaching hardwareAir Flow - Ventilation <strong>System</strong>7.1 - 9.9 cubic metres (250 - 350 cubic ft.) per minute at 5.1 mm (0.2 in.) of water vacuum8 2014-03-01 – J39897


Section 7: Networking <strong>Specifications</strong>Networking <strong>Specifications</strong>The requirements <strong>for</strong> e-Connectivity network connection are located in <strong>the</strong> e-Connectivity® Network Connection<strong>Specifications</strong> <strong>and</strong> Network Form located on-line at:http://apps.orthoclinical.com/notes/contentpages/techdoc.aspx?view=USSPECSThe customer must complete <strong>the</strong> <strong>for</strong>m <strong>and</strong> return it to Ortho-Clinical Diagnostics.ImportantAll requirements <strong>for</strong> lab network connections must be completed be<strong>for</strong>e installation of <strong>the</strong> analyzer(s).2014-03-01 – J39897 9


Section 8: Requirements at Customer <strong>Site</strong>TELEPHONEAir FlowEntrance to <strong>the</strong> room where<strong>the</strong> <strong>System</strong> will be installedA TELEPHONE located so that <strong>the</strong> user can movearound <strong>the</strong> <strong>System</strong> must be available.No air flow within 3.0 m (10 ft) of <strong>the</strong> <strong>System</strong>.Minimum width of 81.3 cm (32 in.)LAB COMPUTER Interface See <strong>the</strong> Interface Specification section of <strong>the</strong>Operator’s Manual.Network Connection Refer to <strong>the</strong> Networking <strong>Specifications</strong> on page 9.Floor Loading 322 kg/m 2 (65 lb./ft 2 )10 2014-03-01 – J39897


Shipping <strong>Specifications</strong>Section 9: Shipping <strong>Specifications</strong>Shipping <strong>Specifications</strong>ReferenceDescriptionA 274.3 cm (108 in.)B 110.5 cm (43.5 in.)C 161.3 cm (63.5 in.)D 215.9 cm (85 in.) *Weight 811.9 kg (1790 lb)* Minimum overhead clearance required <strong>for</strong> unpacking.Detach 42014-03-01 – J39897 11


Section 10: <strong>Specifications</strong> <strong>for</strong> <strong>the</strong> <strong>System</strong> <strong>and</strong> <strong>the</strong> PRINTER<strong>System</strong> <strong>Specifications</strong> Supplied PRINTER <strong>Specifications</strong> ****Reference Description Reference DescriptionA 233.7 cm (92 in.) * J 40.0 cm (15.7 in.)B 83.8 cm (33 in.) K 51.3 cm (20.2 in.)C 132.1 cm (52 in.) L 34 cm (13.4 in.)D 205.7 cm (81 in.) M 61 cm (24 in.)Weight 612.4 kg (1350 lb) Weight 19.7 kg (43.5 lb.)* 264.7 cm (104.2 in.) if AT Accessory is installed. PortPrinter LanguageInstallation <strong>Specifications</strong><strong>for</strong> <strong>the</strong> <strong>System</strong> onlyUSB Version (TBD)Post ScriptInstallation <strong>Specifications</strong><strong>for</strong> <strong>the</strong> Supplied PRINTER ****Reference Description Reference DescriptionE 76.2 cm (2 ft 6 in.) N 17.8 cm (7 in.)F 76.2 cm (2 ft 6 in.) O 45.7 cm (18 in.)G 30.5 cm (12 in.) ** P 61.0 cm (24 in.)H 4.3 m (14 ft. ) *** **** If <strong>the</strong> PRINTER is being supplied by <strong>the</strong>Q 76.2 cm (30 in.)customer, refer to <strong>the</strong> manufacturer’s PRINTERmanual <strong>for</strong> any specifications.** 91 cm (36 in.) if SEISMIC ANCHORAGES, ATAccessory Kit if installed from <strong>the</strong> rear or o<strong>the</strong>rwise30.5 cm (12 in.) if installed from <strong>the</strong> front, or HEATREJECTION PLENUM are to be installed.*** 4.6 m (15 ft.) if AT Accessory is installed.12 2014-03-01 – J39897


<strong>Specifications</strong> <strong>for</strong> <strong>the</strong> 5,1 <strong>and</strong> 4600 HEAT REJECTION PLENUMSection 11: <strong>Specifications</strong> <strong>for</strong> <strong>the</strong> 5,1 <strong>and</strong> 4600 HEAT REJECTIONPLENUMDetach 5HEAT REJECTION PLENUM<strong>Specifications</strong>Reference DescriptionA 107.1 cm (42.2 in.)B 77.5 cm (30.5 in.)C 38.7 cm (15.2 in.)D 17.7 cm (7.0 in.)2014-03-01 – J39897 13


14 2014-03-01 – J39897


Ortho Clinical Diagnostics50 - 100 Holmers Farm WayHigh WycombeBuckinghamshireHP12 4DPUnited KingdomOrtho-Clinical Diagnostics, Inc.100 Indigo Creek DriveRochester, NY 14626