<strong>Zeiss</strong> Axiovert 200 <strong>Microscope</strong> Specs<br />

Filters:<br />

1 blank = LSM???<br />

2 Blue LP FS 02 ex 365/100 (300-400) bs 395 em 420 LP<br />

3 Green FS 10 ex 470/40 (450-490) bs 510 em 540/50 (515-565)<br />

4 Red FS 15 ex 546/12 (534-558) bs 580 em 590 LP<br />

5 DIC<br />

Objectives:<br />

Mag NA Desc coverslip WD (mm)<br />

5x 0.15 C Plan-NEOFLUAR 0.17 13.6<br />

10x 0.3 C Plan-NEOFLUAR 0.17 5.6<br />

20x 0.8 Plan-APOCHROMAT 0.17 0.55<br />

40x 1.3 oil EC Plan-NEOFLUAR 0.17 0.20<br />

63x 1.4 oil Plan-APOCHROMAT 0.17 0.19<br />

40x 1.2 W C-APOCHROMAT c slip adj 0.28<br />

DIC for all except 5x and 10x<br />

Section thickness in um<br />

Wavelength<br />

Obj 488nm 561nm 633mn<br />

5x/0.15 47.2 um 52.7 um 60.1 um<br />

10x/0.3 11.7 13.1 14.9<br />

20x/0.8 1.6 1.7 2.0<br />

40x/1.3 0.9 1.0 1.1<br />

63x/1.4 0.8 0.8 0.9<br />

40x/1.2W 0.9 1.0 1.1<br />

“Optimal” step size is ½ section thickness.