CPAP ModeCPAP mode is selected when SIMV/CPAP is selected and the Breath Rate is set to dashes “- -”. InCPAP mode, when a patient trigger is detected, a patient breath is given. Breaths will be PressureSupport or Spontaneous breaths according to the Pressure Support setting.NOTE<strong>LTV</strong> ® ventilators provide an Apnea Backup mode of ventilation. When the set Apnea Interval(maximum time allowed between the beginning of one breath and the beginning of the next breath) isexceeded, the APNEA alarm is generated and the ventilator will enter Apnea Backup ventilationmode.REMARQUEVentilateur <strong>LTV</strong> ® procurent un mode de ventilation de secours pour l’apnée. Lorsque l’intervalle del’apnée établi (durée maximum allouée entre le début d’une respiration et le début de la respirationsuivante) est excédé, l’alarme APNEA est générée et le ventilateur entre en mode de ventilation desecours pour l’apnée.Operator’s <strong>Manual</strong> <strong>LTV</strong> ® <strong>1200</strong> Ventilator Page 4-3p/n 18247-001, Rev. E