Apnea BackupThe <strong>LTV</strong> ® <strong>1200</strong> ventilator provides an Apnea Backup mode of ventilation. Apnea Backup ventilationbegins when the time since the last breath start is greater than the set Apnea Interval.When an apnea alarm occurs:• If an inspiration is in progress, the ventilator cycles to exhalation.• The ventilator begins Apnea Backup ventilation in the Assist/Control mode according to thecurrent control settings. The active controls are displayed at full intensity, and all other controlsare dimmed.The breath rate for Apnea Backup mode is determined as follows:• If the set Breath Rate is ≥ = 12 bpm, the Apnea breath rate is the set Breath Rate.• If the set Breath Rate is < 12 bpm and the set Breath Rate is not limited by other control settings,the Apnea breath rate is 12 bpm.• If the set Breath Rate is limited to