Conventional <strong>DSC</strong> Cp Measurement<br />

Cp = K x<br />

HFS<br />

– HFMT<br />

Heat Rate x wt<br />

Where:<br />

K = Calibration constant<br />

HFS = <strong>Differential</strong> heat flow<br />

with sample<br />

HFMT = <strong>Differential</strong> heat flow<br />

with empty pans<br />

wt = weight of sample<br />

0<br />

HF<br />

�<br />

endo<br />

Alternative <strong>DSC</strong> Cp Measurement<br />

Cp<br />

=<br />

K x<br />

HF<br />

(HR<br />

HR2<br />

2<br />

– HF<br />

– HR1)<br />

HR1<br />

wt<br />

Where:<br />

K = Calibration constant<br />

HFHR1 = <strong>Differential</strong> heat flow of<br />

sample at HR1 HFHR2 = <strong>Differential</strong> heat flow of<br />

sample at HR2 HR2 = Heating rate 2<br />

= Heating rate 1<br />

HR 1<br />

0<br />

HF<br />

�<br />

endo<br />

HF MT<br />

HF S<br />

Temp.<br />

Temp.<br />

HF HR1<br />

HF HR2