M<strong>DSC</strong> Signals - Nonreversing Heat Flow<br />

(Kinetic Component)<br />

Nonreversing Heat Flow is the kinetic component of the total heat<br />

flow. It is calculated by subtracting the heat capacity component from the<br />

total heat flow using the classical heat flow equation as a theoretical basis.<br />

Nonreversing = Total – Reversing<br />

Basis for Calculation<br />

dH dT<br />

= Cp + f (T, t)<br />

dt dt<br />

dH<br />

= total heat flow<br />

dt<br />

dT<br />

Cp = heat capacity component (reversing)<br />

dt<br />

f (T, t) = kinetic component (nonreversing)<br />

Quench-Cooled PET: Deconvoluted Signals<br />

Heat Flow (W/g)<br />

0.25<br />

0.15<br />

0.05<br />

-0.05<br />

-0.15<br />

-0.25<br />

NONREVERSING<br />

TOTAL<br />

REVERSING<br />

0.25<br />

0.15<br />

0.05<br />

-0.05<br />

-0.15<br />

-0.25<br />

50 100 150 200 250 300<br />

Nonrev Heat Flow (W/g)<br />

0.1<br />

0.0<br />

-0.1<br />

-0.2<br />

Rev Heat Flow (W/g)