Differential Scanning Calorimetry (DSC)
Differential Scanning Calorimetry (DSC)
Differential Scanning Calorimetry (DSC)
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
M<strong>DSC</strong> Signals - Nonreversing Heat Flow<br />
(Kinetic Component)<br />
Nonreversing Heat Flow is the kinetic component of the total heat<br />
flow. It is calculated by subtracting the heat capacity component from the<br />
total heat flow using the classical heat flow equation as a theoretical basis.<br />
Nonreversing = Total – Reversing<br />
Basis for Calculation<br />
dH dT<br />
= Cp + f (T, t)<br />
dt dt<br />
dH<br />
= total heat flow<br />
dt<br />
dT<br />
Cp = heat capacity component (reversing)<br />
dt<br />
f (T, t) = kinetic component (nonreversing)<br />
Quench-Cooled PET: Deconvoluted Signals<br />
Heat Flow (W/g)<br />
0.25<br />
0.15<br />
0.05<br />
-0.05<br />
-0.15<br />
-0.25<br />
NONREVERSING<br />
TOTAL<br />
REVERSING<br />
0.25<br />
0.15<br />
0.05<br />
-0.05<br />
-0.15<br />
-0.25<br />
50 100 150 200 250 300<br />
Nonrev Heat Flow (W/g)<br />
0.1<br />
0.0<br />
-0.1<br />
-0.2<br />
Rev Heat Flow (W/g)