Sandra Hopkins Final Report.pdf - University of Surrey
Sandra Hopkins Final Report.pdf - University of Surrey
Sandra Hopkins Final Report.pdf - University of Surrey
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
70 183 67 60.7 +10.3%102 205 175 159.3 +9.9 %90 566 464 403.5 +15.0 %90 312 222 198.4 +11.9 %81 167 85 78.8 +7.9 %70 182 67 61.7 +8.6 %Table 4 : Measurements to determine the accuracy <strong>of</strong> the Skin Dose Indicator with no added Copper filters in the beamWith no additional Copper filtration in the beam the indicated skin dose is between 8% and 15%higher than the actual skin dose (table 4). Additional measurements carried out at the same kV werefor consistency or were with different amounts <strong>of</strong> attenuation to change the mA and therefore theindicated mGy/min.With various amounts <strong>of</strong> additional copper automatically added into the beam the accuracy <strong>of</strong> theskin dose indicator over reads by 15% to 22% (table 5)DisplayedkVDisplayedmA+/-2%AddedCu bysystemIndicatedmGy/min+/-5%MeasuredmGy/min+/- 4%Differencebetween indicatedand actual83.6 813 0.1 mm 226 191.7 +17.9 %87.2 817 0.1 mm 409 341.6 +19.7 %81.8 809 0.1 mm 338 286.2 +18.1 %86.7 807 0.1 mm 291 255.3 +14.0 %75.6 690 0.2 mm 189 167.2 +13.0 %83.4 681 0.2 mm 180 155.6 +15.7 %64.5 532 0.3 mm 44 38.2 +15.2 %73.5 740 0.3 mm 74 60.5 +22.3 %73.5 471 0.3 mm 60 50.3 +19.3 %64.5 492 0.3 mm 28 22.9 +22.2 %64.5 537 0.3 mm 44 38.4 +14.6 %81 492 0.3 mm 73 60.7 +20.3 %76.1 452 0.6 mm 30 25.5 +17.6 %67.4 646 0.9 mm 18 15.2 +18.4 %73.6 311 0.9 mm 9 7.5 +20 %Table 5 : Measurements to determine the accuracy <strong>of</strong> the Skin Dose Indicator with various Copper filters in the beam21