<strong>Detector</strong> descriptionThe detector is designed for high-performance liquid chromatographyapplications. It can operate as a standalone unit with an integrator or chartrecorder, or with a <strong>Waters</strong> system controller or <strong>Waters</strong> data system.Refer to Chapter B, “Specifications” for system specifications, and toChapter 4, “Preparing Solvents” for solvent considerations.<strong>Waters</strong> <strong>2414</strong> <strong>Refractive</strong> <strong>Index</strong> <strong>Detector</strong>Range and attenuationFeaturesThe detector functions with solvents with refractive indices between 1.00 and1.75. The measurement range of the instrument is 7 × 10 –9 to 5 × 10 –4refractive index units full scale (RIU-FS mode).The detector has these key features:• Standalone programmability – Stores up to 10 user-defined programs (ormethods) consisting of up to 16, programmable, timed events and 1threshold event (in standalone RIU mode).<strong>Detector</strong> description 1-11