Note: When the detector is operating under 410 mode, the method icon is notpresent or if under control by a system controller of a data system, the remoteicon is displayed.The detector method includes all parameters that are necessary to reproducea chromatographic separation.Timed eventsYou can program up to 16 timed events to the nearest 0.01 minute. As youenter timed events, each new event is appended to the end of the timed eventlist. If you enter a time that is not in sequence with the events enteredpreviously, the timed event list is sorted automatically when you press Next.The <strong>2414</strong> detector allows programming of the timed events shown in thefollowing table.Timed event parameterNumber Event Units Range Default1 AutoZero Does not apply Does not apply Disabled2 Polarity Does not apply 1. – +2. +3 RIU-FS µRIU 1 to 500 Disabled4 Chart mark Does not apply Does not apply Disabled5 Start Purge Does not apply Does not apply Disabled6 End Purge Does not apply Does not apply Disabled7 Switch µRIU 1. On Disabled2. Off3. Pulse4. RectangularWave5. Toggle8 Threshold µRIU 0.001 to Disabled500.0009 Start Recycle Does not apply Does not apply Disabled10 End Recycle Does not apply Does not apply Disabled5-18 Operating the <strong>Detector</strong>