3.4 Commands and responses <strong>MT</strong>-<strong>SICS</strong> level 3 (subset for HalogenMoisture Analyzer HB43)All Halogen Moisture Analyzer specified commands are combined in <strong>MT</strong>-<strong>SICS</strong> level 3 forHalogen Moisture Analyzers HR73, HG53 and HB43.HA01Reset application / escapeCommand HA01 Reset application / escapeResponse HA01/A Application resetCommentThis command has the same effect as the Reset key, see operating instructions of theHalogen Moisture Analyzer. It terminates all current commands and activities.HA02Set factory settingsCommand HA02 Set factory setting of the menu and methodparametersResponse HA02/A Mean and method parameters set to factory settingCommentAll menu parameters are reset to factory settings except RS interface settings and language.This command terminates a drying.HA03Switch keypad on/offCommand HA03/x x = 0 Keypad of Halogen Moisture Analyzerswitched offx = 1 Keypad of Halogen Moisture Analyzerswitched on (factory setting)Response HA03/A Command executedHA03/L Command understood, parameter wrong26 METTLER TOLEDO <strong>MT</strong>-<strong>SICS</strong> HB43 Reference Manual