HA21Inquiry of heating module positionCommand HA21 Inquiry of heading module positionResponse HA21/A/x x = 0 Heating module closedx = 1 Heating module openHA22Inquiry of last balance adjustmentCommand HA22 Inquiry of last successful balance adjustmentResponse HA22/A/x1/x2/x3/x4/x5/x6x1 Number of the successful adjustmentsx2 Day of the last successful adjustmentx3 Month of the last successful adjustmentx4 Year of the last successful adjustmentx5 Hour of the last successful adjustmentx6 Minute of the last successful adjustmentExampleCommandResponseHA22HA22/A/15/02/04/2000/09/34A total of 15 successful balance adjustments havebeen performed. The last took place on April 02, 2000at 9.34.Comments• The time of the last successful balance adjustment is specified in the 24-hour format.• Possible years are 1999 … 2099.• The counter for the balance adjustments runs to 65535.30 METTLER TOLEDO <strong>MT</strong>-<strong>SICS</strong> HB43 Reference Manual