Example 2Command HA26/2 Inquiry of drying dataResponseHA26/A/1/2/2.672/2.467/92.33/143Drying is running, result requested in % dry content,wet weight 2.672 g, dry weight 2.467 g, 92.33 %dry content, drying for 143 seconds in progressComment• If a drying is inexistent (e.g. after a RAM LOST), the parameters x3..x6 are set to 0.• If the measuring results exceed the tolerances for ATRO result display (L-999.99 % AMor >999.99 % AD) the selected results in x2 = 4 AM or x2 = 5 AD will automatically betransferred in x2 = 3 MC or x2 = 2 DC respectively.HA27Inquiry of drying resultCommand HA27/x1 Inquiry of drying data in configurable display modex1 = 0 currently set display modex1 = 1 Gramsx1 = 2 DC (dry content)x1 = 3 MC (moisture content), (factory setting)x1 = 4 AM (ATRO moisture content)x1 = 5 AD (ATRO dry content)Response HA27/A/x1/x2x1 Drying status (always 7 digit number)x2 Display mode (g, %DC, %MC, %AM, %AD)HA27/IResponse not available (drying in progress)ExampleCommand HA27/3 Inquiry of drying resultResponse HA27/A////-73.25%MCDrying result -73.25 % MCCommentIf the measuring results exceed the tolerances for ATRO result display (L-999.99 % AM or> 999.99 % AD) the selected results in x1 = 4 AM or x1 = 5 AD will automatically betransferred in x1 = 3 MC or x1 = 2 DC respectively.34 METTLER TOLEDO <strong>MT</strong>-<strong>SICS</strong> HB43 Reference Manual