Setting print intervalCommand HA62/x1/x2/ … /x5Set method parameters regarding print intervalx1 … x5 represent the individual parameters (seebelow)Response HA62/A Method parameter setHA62/LCommand understood, parameter wrongParameters x1 Number of the methodPossible values for HB43 1x2 Target weight in gramsPossible settings 0x3 Print intervalx3 = 1 No print interval set, manual initiation ofprintoutx3 = 4 Printout every 30 secondsx3 = 5 Printout every 60 secondsx3 = 9 Printout every 300 secondsx4 Method name""x5 Code" "ExampleCommandHA62/1/0/5/""/""Print interval set to 60 secondsResponse HA62/A Method parameter setComments• Setting the method parameters terminates a drying.• The additional parameters are implemented in the HR/HG Halogen Moisture Analyzers.The above settings assure direct transfereability of control programs.40 METTLER TOLEDO <strong>MT</strong>-<strong>SICS</strong> HB43 Reference Manual