oxygen towards the 3 mm tip has a measurable effect on theresponse time. For optimum operation conditions it is thereforere<strong>com</strong>mended to utilize these robust probes in permanently stirredliquids (this diffusion limitation is negligible in gaseous samples;see also the respective sensor specifications).Ensure that the <strong>com</strong>plete sensing surface of the Robust OxygenProbe is covered <strong>com</strong>pletely by the sample and preventmechanical stress (e.g. scratching) to the sensing surface!5.4 Sensor Spots<strong>Pyro</strong>Science offers arange of contactlessoxygen sensors,<strong>com</strong>prising sensorspots, respirationvials and flowthroughcells. Thesensor spots arecoated with theREDFLASH indicatorwith PET foil or glassas carrier material.The sensor spots allow oxygen measurements in closed vessels in<strong>com</strong>bination with the <strong>FireStingO2</strong> equipped with Optical Fibers(e.g. item no. SPFIB or SPFIB-CL2). The sensor spots havestandard diameters of 5 mm (item no. OXSP5) and can beoptionally obtained with an optical isolation (black coating) on thesensing surface (e.g. item no. OXSP5OI, OXSG8OI). Opticalisolation is re<strong>com</strong>mended if strong illuminationmight disturb the measurement or if no light shallbe introduced into the sample.20