7 Calibration with “<strong>Pyro</strong> Oxygen Logger”This chapter describes the possible calibration modes for oxygensensors using the logger software “<strong>Pyro</strong> Oxygen Logger”. Pleasenote that the possible calibration modes differ depending on theconnected <strong>FireStingO2</strong> device:If your device has integrated humidity and pressure sensors (e.g.<strong>FireStingO2</strong> devices with micro-USB connector, firmware >3.0)proceed with chapter 7.1.If your device does not possess integrated humidity and pressuresensors (e.g. <strong>FireStingO2</strong> devices with classical full-size USBconnector, firmware 2.30), then proceed with chapter 7.2.For many calibration modes, the atmospheric pressure and therelative humidity of the ambient air are important parametersensuring a high precision of the calibration. Formerly it was on theuser´s authority to measure and adjust manually these parameters.But the <strong>FireStingO2</strong> devices <strong>com</strong>e with integrated pressure andhumidity sensors, which measure these parameters nowautomatically, ensuring a straightforward calibration procedureand eliminating many possible error sources.This chapter covers only the necessary steps during the calibrationprocedure. For details regarding the preparation of the calibrationstandards refer to chapter 8.48