Table 8 shows the results for a duplicate analysis test<br />

with the LCS sample. Agreement between duplicates is<br />

generally within 1%, well within the EPA limit of 20%.<br />

ELEMENT ANALYSIS 1 ANALYSIS 2 % DIFFERENCE<br />

As 23.1µg/g 23.3µg/g 0.8<br />

Ba 407µg/g 403µg/g 1.0<br />

Ca 2.55% 2.54% 0.4<br />

Cd 3.50µg/g 3.46µg/g 1.1<br />

Co 14.3µg/g 14.4µg/g 0.7<br />

Cr 132µg/g 131µg/g 0.8<br />

Cu 100.2µg/g 101µg/g 0.8<br />

Fe 4.03% 4.04% 0.3<br />

K 2.02% 2.03% 0.6<br />

Mg 1.21% 1.20% 0.3<br />

Mn 547µg/g 543µg/g 0.8<br />

Na 0.53% 0.54% 1.8<br />

Ni 43.8µg/g 44.1µg/g 0.7<br />

Pb 163µg/g 163µg/g 0.2<br />

Sb 3.61µg/g 3.65µg/g 1.1<br />

Ti 0.46% 0.46% 0.4<br />

V 93.0µg/g 92.3µg/g 0.8<br />

Zn 440µg/g 435µg/g 1.1<br />

Table 8. NBS 2704 Buffalo River Sediment Duplicate Analysis Test<br />

Table 9 shows the recoveries from spiked samples of<br />

the LCS, again all well within the EPA 25% limit.<br />

ELEMENT CERTIFIED VALUE SPIKE CONC MEASURED VALUE % REC<br />

www.thermo.com/elemental<br />

µg/L µg/L µg/L<br />

As 230 50 276 92<br />

Ba 4140 100 4234 94<br />

Cd 34.5 10 43.6 91<br />

Co 140 10 151 110<br />

Cr 1350 50 1402 104<br />

Cu 986 50 1033 94<br />

Mn 5550 100 5652 102<br />

Ni 441 10 450 90<br />

Pb 1610 50 1656 92<br />

Sb 37.9 10 47.0 91<br />

V 950 50 1004 108<br />

Zn 4380 100 4482 102<br />

Table 9. NBS 2704 Buffalo River Sediment Spike Sample Recovery Data<br />

In order to check that the method works for different<br />

sample matrices, a reference sample of a sewage sludge<br />

(BCR146) was analysed using exactly the same procedure.<br />

Table 10 compares the measured and certified results for<br />

the sediment sample, demonstrating the excellent<br />

agreement. Values in brackets are advisory figures only.<br />

USA: 5225 Verona Road<br />

Madison, WI 53711-4495<br />

Tel: +1 800 201 8132, +1 608 276 6100<br />

Fax: +1 608 273 5046<br />

Email: spectroscopy@thermo.com<br />

ELEMENT MEASURED CONCENTRATION CERTIFIED VALUE<br />

Ca 10.2% (10.14%)<br />

Cd 75.8µg/g 77.7 +/- 2.6<br />

Co 11.9µg/g 11.8 +/- 0.7<br />

Cu 928µg/g 934 +/- 24<br />

Fe 1.80% (1.86%)<br />

K 0.49% (0.48%)<br />

Mg 2.01% (1.99%)<br />

Mn 570µg/g 588 +/- 24<br />

Na 0.21% (0.22%)<br />

Ni 276µg/g 280 +/- 18<br />

P 2.52% (2.58%)<br />

Pb 1260µg/g 1270 +/- 28<br />

Zn 4040µg/g 4059 +/- 90<br />

Table 10. BCR146 Sewage Sludge Data<br />

5. Conclusions<br />

The IRIS Intrepid II demonstrates SW-846 Method 6010B<br />

compliant analysis for a range of sample types and easily<br />

copes with the stringent interference checks and AQC<br />

requirements stipulated for use in this method.<br />

The IRIS Intrepid II is the latest generation of IRIS<br />

ICP Spectrometers from Thermo Electron. It incorporates<br />

a new design of the exclusive charge injection device<br />

(CID) detector housed in a highly regulated optical<br />

system. The result of these product developments is an<br />

instrument that is capable of providing extremely low<br />

detection limits and excellent long term stability. This<br />

makes the IRIS Intrepid II ideally suited for cost-effective<br />

elemental analysis in the environmental field.<br />

References<br />

USEPA SW-846 Method 3050B, Revision 2, December 1996.<br />

USEPA SW-846 Method 3051A, Revision 1, January 1998.<br />

USEPA SW-846 Method 6010B, Revision 2, December 1996.<br />

USEPA SW-846 Method 6010C, Revision 3, November 2000.<br />

UK: 19 Mercers Row<br />

Cambridge, CB5 8BZ, UK<br />

Tel: +44 (0)1223 345410<br />

Fax: +44 (0)1223 345411<br />

Email: spectroscopy.uk@thermo.com<br />

Switzerland: En Vallaire Ouest C<br />

case postale, CH-1024, Ecublens<br />

Tel: +41 (0)21 694 71 11<br />

Fax: +41 (0)21 694 71 12<br />

Email: spectroscopy.ch@thermo.com<br />

Thermo Electron Corporation<br />

has direct subsidiary offices in<br />

North America, Europe and<br />

Japan. To complement these<br />

direct subsidiaries, we maintain<br />

a network of representative<br />

organizations throughout the<br />

world. Use this reference list or<br />

visit our Web site to locate the<br />

representative nearest you.<br />

Australia<br />

Tel: +61 (0)2 9898 1244<br />

Austria<br />

Tel: +43 (0)1 333 50 34 0<br />

Belgium<br />

Tel: +32 2482 3030<br />

Canada<br />

Tel: +1 905 890 1034<br />

China<br />

Thermo ARL:<br />

Tel: +86 10 6833 6715<br />

Thermo Elemental:<br />

Tel: +86 10 6592 0232<br />

Thermo Nicolet:<br />

Tel: +86 10 6597 3388 ext 2912<br />

France<br />

Tel: +33 (0)1 39 30 53 00<br />

Germany<br />

Tel: +49 (0)6102 3671 0<br />

Italy<br />

Tel: +39 02 6601 6351<br />

Japan<br />

Thermo Elemental:<br />

Tel: +81 774 201245<br />

Thermo Nicolet:<br />

Tel: +81 45 450 1112<br />

Netherlands<br />

Tel: +31 76 5724840<br />

Nordic<br />

Thermo Nicolet:<br />

Tel: +358 9 3291 00<br />

Thermo ARL:<br />

Tel: +46 (0)8 556 468 31<br />

South Africa<br />

Tel: +27 (0) 11 570 1840<br />

Spain<br />

Tel: +34 (91) 657 49 30<br />

©2003 Thermo Electron Corp.<br />

All rights reserved worldwide.<br />

We make no warranties,<br />

expressed or implied, in<br />

this product summary, and<br />

information is subject to<br />

change without notice.<br />

All product and company<br />

names are property of their<br />

respective owners.<br />

AN_E0637 06/03

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!