note<br />
7.2. Preparation of Test St<strong>and</strong>ard Solution<br />
<strong>Addendum</strong> <strong>to</strong> <strong>PAL</strong> <strong>System</strong> <strong>User</strong> <strong>Manual</strong><br />
<strong>PAL</strong> ITEX-2 Option<br />
The test or calibration st<strong>and</strong>ard solution as received (concentration<br />
100 µg/mL of each compound) is diluted 1:10 with methanol.<br />
This secondary st<strong>and</strong>ard solution is used for the tests as described below.<br />
Ensure that the methanol used for the dilution step is of high purity.<br />
7.2.1. Selective GC Detec<strong>to</strong>r, e.g. MSD <strong>and</strong> ECD<br />
Add 1 µL of the secondary st<strong>and</strong>ard solution directly in<strong>to</strong> a 20 mL Headspace<br />
vial <strong>and</strong> seal the vial immediately. Use a microcap device or a GC syringe <strong>to</strong><br />
dose the aliquot of the st<strong>and</strong>ard solution.<br />
• Secondary st<strong>and</strong>ard solution: Concentration 10 µg/mL<br />
• 1 µL aliquot = 10 ng/ µL<br />
7.2.2. GC Detec<strong>to</strong>r FID, TCD or any other Non-selective Device<br />
Add 10 µL of the secondary st<strong>and</strong>ard solution directly in<strong>to</strong> a 20 mL<br />
Headspace vial <strong>and</strong> seal the vial immediately. Use a microcap device<br />
or a GC syringe <strong>to</strong> dose the aliquot of the st<strong>and</strong>ard solution.<br />
• Secondary st<strong>and</strong>ard solution: Concentration 10 µg/mL<br />
• 10 µL aliquot = 100 ng/ µL<br />
If the calibration st<strong>and</strong>ard is used only for the less sensitive GC Detec<strong>to</strong>rs,<br />
such as FID or TCD, then the primary st<strong>and</strong>ard with a concentration of<br />
100 µg/mL can be used directly. Add a 1 µL aliquot of the primary st<strong>and</strong>ard<br />
solution in<strong>to</strong> a 20 mL headspace vial.<br />
7.3. Instrument Preparation for the ITEX Test<br />
7.3.1. Cycle Composer Macros or ICC- Cycles<br />
The Cycle Composer macros or ICC-cycles used for the ITEX-2 tests are:<br />
• Static Headspace: ITEX2_TestHS<br />
• ITEX enrichment: ITEX2_TestITEX<br />
7.3.2. GC Method Parameters<br />
• GC column: nearly any capillary column can be used <strong>to</strong> separate the<br />
compounds in the test sample<br />
• GC Injec<strong>to</strong>r temperature: 200 <strong>to</strong> 250°C, split less for 1 min.<br />
• Detec<strong>to</strong>r: settings depend on the detec<strong>to</strong>r used<br />
• GC oven temperature: 40 0 C, hold 1 minute, 10 0 C/min up <strong>to</strong> 150 0 C,<br />
hold 2 minutes<br />
36