E. <strong>PAL</strong> ITEX-2 Option <strong>Installation</strong><br />
1. General <strong>System</strong> Overview<br />
Figure 1. <strong>PAL</strong> ITEX-2 Option.<br />
<strong>Addendum</strong> <strong>to</strong> <strong>PAL</strong> <strong>System</strong> <strong>User</strong> <strong>Manual</strong><br />
<strong>PAL</strong> ITEX-2 Option<br />
The <strong>PAL</strong> ITEX-2 Option is an add-on module for the <strong>PAL</strong> <strong>and</strong> <strong>PAL</strong>-xt <strong>System</strong>s<br />
<strong>to</strong> extend the range of headspace techniques. ITEX can be considered an<br />
au<strong>to</strong>mated dynamic headspace technique. In classical headspace technique,<br />
one sampling is done per vial. In this device, analytes are adsorbed (loaded)<br />
on a sorbent trap contained within a side-hole needle attached <strong>to</strong> a gas-tight<br />
syringe. Following this loading procedure, analytes are thermally desorbed<br />
<strong>and</strong> led directly in<strong>to</strong> the GC injection port, then analyzed by the required<br />
technique, mainly using GC-MS detection.<br />
Loading takes place by continuously moving the headspace syringe plunger<br />
up <strong>and</strong> down, passing the headspace gas phase that exists above a sample<br />
through the sorbent trap inside the needle numerous times.<br />
Dynamic headspace extraction increases the concentration of analyte in order<br />
<strong>to</strong> extend the range of detection as compared <strong>to</strong> conventional static<br />
headspace techniques. Required levels of detection can be reached by<br />
applying more pumping (up <strong>and</strong> down) cycles. To reach the detection limit in<br />
the range of a ppm <strong>to</strong> ppb concentration, three <strong>to</strong> twenty strokes are typically<br />
applied <strong>and</strong>, for the ppt-level of concentration, 40 <strong>to</strong> 60 pumping strokes may<br />
be necessary.<br />
The <strong>PAL</strong> ITEX-2 Option consists of a dedicated gas-tight syringe with an<br />
adapter for a replaceable side-hole needle <strong>and</strong> a heated syringe cartridge.<br />
The side-hole needle includes an integrated trap (st<strong>and</strong>ard trap material is<br />
Tenax TA) <strong>and</strong> a trap/desorber heater.<br />
8