<strong>ME5000</strong> Test Measurements<br />

4. FREQUENCY MEASUREMENTS<br />

Since <strong>the</strong> accuracy <strong>of</strong> any EDM instrument is at least partly dependent upon <strong>the</strong> accurate knowledge<br />

<strong>of</strong> <strong>the</strong> modulation frequency, it is important to ensure that <strong>the</strong>re is no significant frequency drift over time.<br />

<strong>The</strong> basic equation for dete mining <strong>the</strong> distance is shown in (3). <strong>The</strong> distance D equals a whole number m<br />

<strong>of</strong> half-wavelengths U2, plus some fiactionf<strong>of</strong> an additional half-wavelength.<br />

k x<br />

D= mpfz (3)<br />

Three different methods are used in EDM instruments in order to fmd f, <strong>the</strong> fractional part <strong>of</strong> a half-<br />

wavelength.<br />

1. I&a-red instruments, such as <strong>the</strong> DM503, use a fued t?equency <strong>and</strong> measure <strong>the</strong> phase<br />

difference between <strong>the</strong> reference signal <strong>and</strong> <strong>the</strong> returning beam. Such an approach, though <strong>the</strong><br />

simplest <strong>and</strong> cheapest, introduces cyclic error.<br />

2. <strong>The</strong> frequency <strong>of</strong> <strong>the</strong> ME3000 <strong>Mekometer</strong> is allowed to vary as changing atmospheric conditions<br />

change <strong>the</strong> dimensions <strong>of</strong> <strong>the</strong> reference cavity. A system <strong>of</strong> prisms is moved within <strong>the</strong><br />

instrument so as to increase <strong>the</strong> path length until <strong>the</strong> fractional component a is eliminated. This<br />

approach eliminates cyclic error.<br />

3. <strong>The</strong> <strong>ME5000</strong> <strong>Mekometer</strong> varies <strong>the</strong> frequency until <strong>the</strong>re is an integral number <strong>of</strong> wavelengths in<br />

<strong>the</strong> distance. Like <strong>the</strong> ME3000, this approach eliminates cyclic error.<br />

-<strong>The</strong> frequency cahbration <strong>of</strong> <strong>the</strong> <strong>ME5000</strong> is more straightforward than that <strong>of</strong> <strong>the</strong> earlier ME3000<br />

<strong>Mekometer</strong>. Because <strong>the</strong> carrier wave <strong>of</strong> <strong>the</strong> <strong>ME5000</strong>, a HeNe laser (h = 632.8 nm), is continuously<br />

modulated <strong>the</strong> modulation frequency can be directly measured by any frequency counter capable <strong>of</strong><br />

operating at 500 MHz (Fig. 2). Since <strong>the</strong>re is no reference cavity in <strong>the</strong> <strong>ME5000</strong>, <strong>the</strong> modulation<br />

frequency is not dependent on <strong>the</strong> ambient temperature <strong>and</strong> pressure as is <strong>the</strong> ME3000.<br />

A SMA jack, similar to that on <strong>the</strong> ME3000, is located on <strong>the</strong> underside <strong>of</strong> <strong>the</strong> <strong>ME5000</strong>. A SMA-<br />

BNC adapter is included .with <strong>the</strong> instrument so that a st<strong>and</strong>ard c0axia.l cable can be connected to this<br />

outlet. <strong>The</strong> function switch should be set to ei<strong>the</strong>r Battery or Remote. Set to Battery, <strong>the</strong> frequency<br />

currently set is shown on <strong>the</strong> display. Set to Remote, any frequency within <strong>the</strong> range <strong>of</strong> <strong>the</strong> instrument<br />

(460 - 5 10 MHz) can be set externally.<br />

SMA-BNC<br />

Adapter<br />

FIGURE 2. Schematic for <strong>ME5000</strong> frequency calibration<br />

600 MHz<br />

Frequency<br />

counter<br />

At <strong>SLAC</strong> we use a HP 5342A microwave frequency counter in one <strong>of</strong> <strong>the</strong> Klystron test labs. It is<br />

calibrated annually <strong>and</strong> is reliable to a few tens <strong>of</strong> Hertz. As long as power is supplied to <strong>the</strong> counter, <strong>the</strong><br />

crystal is kept at a stable temperature even when <strong>the</strong> main power switch is <strong>of</strong>f.<br />