<strong>ME5000</strong> Results<br />
Tme from power on (hmm)<br />
04 4<br />
460 466 470 475 480 485 450 4% Sal SC5 510<br />
Froquoncy (MHz)<br />
FIGURE 2. Change <strong>of</strong> characteristic curve with time<br />
TABLE 2. Change <strong>of</strong> modulator b<strong>and</strong>width with time: instrument 357089 (SL4C)<br />
1 Time 1<br />
h:mm<br />
OiKl<br />
0:20<br />
1:OO‘<br />
t 2:oo<br />
Normal Range (MHz) I Extended Range (MHz) I<br />
1.2. Modulator B<strong>and</strong>width<br />
<strong>The</strong> oominal b<strong>and</strong>widths are 15 MHz for normal <strong>and</strong> 30 MHz for extended modes. Measurements<br />
show that <strong>the</strong> normal b<strong>and</strong>width is always ei<strong>the</strong>r 14.130 or 14.803 (<strong>the</strong> difference between <strong>the</strong>se two<br />
values is <strong>the</strong> spacing between measurements during <strong>the</strong> modulator calibration in which <strong>the</strong> b<strong>and</strong>width is<br />
determined). <strong>The</strong> extended b<strong>and</strong>width, by contrast, varies considerably between instruments, Table 3.<br />
TABLE 3. Modulator B<strong>and</strong>widths<br />
1rLstmment Nonnal Range (MHz) Extended Range (MHz)<br />
Lob s/n Date Lower Upper Range Lower Upper Raqe<br />
LBL 357037 01-06-92 472.112 486.915 14.803 465.383 494.644 28.261<br />
Fennilab 357046 01-13-92 476.149 490.952 14.803 474.130 492.971 18.841<br />
CEBAF 357036 01-27-92 476.822 490.952 14.130 473.458 494.317 20.859<br />
Argonne 357086 02-18-92 474.130 488.934 14.804 468.747 494.317 25.570<br />
Brookhaven 357088 03-1692 476.149 490.280 14.131 472.785 493.644 20.859<br />
<strong>SLAC</strong> 357089 06-10-92 472.112 486.242 14.130 464.710 493.644 28.934<br />
ssc 357075 08-03-92 473.458 487.588 14.130 470.766 490.280 19.513<br />
ssc 357081 08-06-92 472.785 486.915 14.130 469.420 490.280 20.859<br />
1.3. Scale Factor<br />
<strong>The</strong> frequency <strong>of</strong> each instrument is measured by <strong>Kern</strong> (L&a) in Switzerl<strong>and</strong> prior to shipping, <strong>and</strong><br />
tbe result is shipped with <strong>the</strong> instrument in <strong>the</strong> form <strong>of</strong> a calibration certificate. This calibration consists<br />
<strong>of</strong> a single measurement at c. 480 MHz. At <strong>SLAC</strong> a more comprehensive set <strong>of</strong> frequency measurements<br />