Variance Component Analysis<br />

2. Example output file for VARCOM<br />

DATE : 07-14-89. INPUT-FILENAME : VARCOM<br />

TIME : 14:56 FILETYPE : INPUT<br />

PROGRAM VARCOM<br />

TEST EXAMPLE FRiiHLICH MEKOMETER MEAS. (D2) MUNICH<br />

VARIANCE COMPONENT ESTIMATION<br />

___--------------------------<br />

_----------------------------<br />

COMPUTATION WITH ADDITION CONSTANT<br />

_----------------___--------------<br />

NUMBER OF BASELINE POINTS = 7<br />

ALPHA1 = 1.00000 ALPHA2 = 0.01000 H = 1.00<br />

M(s)**2 = SIGMA1**2 + SIGMA2**2 * S**(2*H)<br />

*************************************************************<br />

RESULTS AFTER 11 ITERATIONS<br />

MO f 1.000‘<br />

SIGMA1**2 = 0.023 (mm) M(SIG1**2) = 0.022 (mm)<br />

SIGMA2**2 = 0.310 otun/Km) M(SIG2**2) = 0.313 (m/Km)<br />

FROM TO SO (mf dS bnm) SO + dS R.M.S.(mm)<br />

-----~~~--~_---------~~~~~~--~-----------~-~~~~~---------~~~-~~~<br />

------~--~-----------~~~~~~~---~------~~--~~~~~~~------~~--~~-~~<br />

12 26.50860 -0.517 26.50808 0.115<br />

13 161.51630 -0.850 161.51545 0.129<br />

14 .243.01060 -0.543 243.01006 0.150<br />

15 431.98070 -1.171 431.97953 0.169<br />

16 485.52540 -0.839 485.52456 0.197<br />

17 540.01610 -0.673 540.01543 0.233<br />

ADDITION CONSTANT C = -0.702 (mm) M(c) = 0.087<br />