Adjusting the probe (see also chapter 6):<br />
English<br />
The distance between the<br />
MiniCoder surface and the<br />
toothed wheel and reference<br />
mark (gap d) must not exceed<br />
the tolerances stated in the<br />
table (take into account a possible eccentricity of the toothed<br />
wheel).<br />
In this connection, please note that the gap may alter critically<br />
at large temperature fluctuations and different thermal<br />
expansion coefficients of the toothed wheel and the carrier<br />
plate.<br />
➤ If distance is too wide the measuring accuracy is affected negatively (signal amplitudes<br />
are possibly too small).<br />
➤ If distance is too small saturation might occur possibly leading to pulse losses.<br />
Dimensioned<br />
drawing:<br />
Module<br />
Gap d<br />
measure<br />
Tolerance<br />
0.3 0.10 mm ± 0.02 mm<br />
0.5 0.15 mm ± 0.03 mm<br />
measures<br />
in mm<br />
<strong>GEL</strong> <strong>244</strong> 17