28<br />

IQmark Digital Spirometer Operation Manual<br />

Clicking on the handle will display the following calibration screen.<br />

Calibration Screen<br />

Check to see if Sensor: Ready is displayed in the upper right hand corner of the graph<br />

window. If you are using a Spirometer with a COM connector (DB9) with a model/serial<br />

number beginning with 51xxxx, you must place the DPM in the handle to turn it on.<br />

Please note that the handle will remain on whenever the DPM is inserted. Be sure to<br />

remove the DPM after calibration or testing to conserve the life of your batteries.<br />

Barometric Pressure<br />

The Pressure must be set prior to the first time you calibrate the system. It must be<br />

changed only if you change the location of the instrument and the altitude changes. Click<br />

the Settings button to open Spirometry Settings. Select the Calibration tab and enter<br />

the Barometric Pressure in the Default Barometric Pressure box. If you do not know the<br />

barometric pressure, you can select Calculate Barometric Pressure From Altitude and<br />

enter your altitude. The software will automatically calculate and store the usual<br />

barometric pressure for that altitude.<br />

NOTE: The proper Barometric Pressure or altitude MUST be entered<br />

to assure a proper calibration.