05.05.2013 Views

OPERATION / SERVICE MANUAL - Midmark

OPERATION / SERVICE MANUAL - Midmark

OPERATION / SERVICE MANUAL - Midmark

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

30<br />

IQmark Digital Spirometer Operation Manual<br />

Calibration Pump Window<br />

Follow the instructions displayed. Use the red dotted line and/or the blue timer arrow as a<br />

target flow rate. Try to inject the air from the syringe at a rate close to that of the red line.<br />

Remember that the syringe handle attached to the plunger must be pulled all the way out<br />

before you inject. Once the handle is all the way out, inject the air from the syringe until<br />

the handle is all the way in. PUSH THE HANDLE IN AND DRAW THE HANDLE OUT<br />

SMOOTHLY – TRY NOT TO “BANG” THE PLUNGER AS YOU PERFORM THE<br />

MANEUVER. “Banging” the plunger can cause the timer arrow to start to count. If it<br />

does, just wait for the timer to reset and inject the plunger. This will not affect the<br />

calibration results.<br />

If the pump is performed correctly, the program will prompt you to begin a second<br />

injection. Repeat the process, injecting all of the air in the syringe. If the injection is not<br />

performed correctly, the system will discard the attempt and prompt you to inject again.<br />

After three correctly performed injections, the system will automatically calculate a<br />

correction factor and prompt you to perform a verification pump. ALWAYS VERIFY<br />

YOUR CALIBRATION. After a verification injection, the software will display the<br />

measured volume and the percentage difference from 3.0 Liters. The ATS recommends<br />

that the verified volume should be between 2.91 and 3.09 (+/- 3.0 %) to be accepted. Be<br />

sure to check these numbers before accepting the verification.<br />

The verification acceptance dialog box will display those numbers for quick and easy<br />

comparison. Click Yes if the verification flow is within the recommended parameters. If<br />

the flow is not within the parameters, click No and repeat the verification flow.<br />

Verification Acceptance Dialog Box<br />

After you accept the verification, the system automatically saves the calibration. To print<br />

the calibration report, click Print and select PRINT from the menu. To view the<br />

calibration report, click Print and select PRINT PREVIEW from the menu.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!