30<br />
2. You will be prompted to confirm your intention with the<br />
display Confirm Clr Press Enter. Press ↵.<br />
3. The display will show Memory Cleared for a few seconds<br />
and will go back to the Setup Menu.<br />
Log Interval<br />
The log interval is the time interval used between recorded data<br />
points during Run <strong>Manual</strong> logging operations (log intervals for<br />
Prog-1 and Prog-2 are set using TRAKPRO TM Software). For<br />
example, if the log interval is set to 30 seconds, a data point is<br />
stored to memory once every 30 seconds. One reason you might<br />
want to use a longer log interval is to conserve memory.<br />
The SIDEPAK AM510 always makes a measurement once every<br />
second regardless of the log interval selected. Using a log interval<br />
greater than one second results in stored data points that are<br />
averages of the 1-second readings. For example, a 10-second log<br />
interval will result in one stored data point every 10 seconds. Each<br />
of those data points will represent the computed average of ten, 1second<br />
readings.<br />
Do not confuse log intervals with time constants. Log intervals only<br />
affect recorded readings. Time constants only affect the AM510<br />
LCD display.<br />
There are five preprogrammed choices for log intervals. If<br />
necessary, the list of available log intervals may be permanently<br />
altered using TRAKPRO TM Software.<br />
After selecting Log Interval from the Setup Menu, the display will<br />
show LOG INTVLx where "x" is a number from 1 to 5. This is the<br />
currently selected log interval. The value of that log interval is<br />
shown on the second line.<br />
Use the ▲ ▼ keys to scroll through the available log intervals.<br />
Press ↵ when the interval you wish to use is displayed. The factorypreset<br />
choices are:<br />
• LOG INTVL1: 1 sec<br />
• LOG INTVL2: 10 sec<br />
• LOG INTVL3: 30 sec<br />
• LOG INTVL4: 1 min<br />
• LOG INTVL5: 5 min<br />
Chapter 3